Towards a Customizable, SLA 3D-Printed Biliary Stent: Optimizing a Commercially Available Resin and Predicting Stent Behavior with Accurate In Silico Testing.

Victoria Cordista, Sagar Patel, Rebecca Lawson, Gunhee Lee, Morgan Verheyen, Ainsley Westbrook, Nathan Shelton, Prakriti Sapkota, Isabella Zabala Valencia, Cynthia Gaddam and 1 more

Abstract read
In one paragraph

Article in Polymers, 2024. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 3 papers.

0numbers the graph read from it
0cells of the map it votes in
3citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

3 citing papers in PubMed.

  1. Review
  2. Review
  3. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

11 authors.

Victoria CordistaSchool of Engineering, Mercer University, Macon, GA 31207, USA.
Sagar PatelSchool of Engineering, Mercer University, Macon, GA 31207, USA.
Rebecca LawsonSchool of Engineering, Mercer University, Macon, GA 31207, USA.
Gunhee LeeSchool of Engineering, Mercer University, Macon, GA 31207, USA.
Morgan VerheyenSchool of Engineering, Mercer University, Macon, GA 31207, USA.
Ainsley WestbrookSchool of Engineering, Mercer University, Macon, GA 31207, USA.
Nathan SheltonSchool of Engineering, Mercer University, Macon, GA 31207, USA.
Prakriti SapkotaSchool of Engineering, Mercer University, Macon, GA 31207, USA.ORCID 0009-0005-6007-2203
Isabella Zabala ValenciaSchool of Engineering, Mercer University, Macon, GA 31207, USA.ORCID 0009-0009-0422-5104
Cynthia GaddamSchool of Engineering, Mercer University, Macon, GA 31207, USA.ORCID 0009-0001-9949-6485
Joanna ThomasSchool of Engineering, Mercer University, Macon, GA 31207, USA.ORCID 0000-0002-1358-3167

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

Inflammation of the bile ducts and surrounding tissues can impede bile flow from the liver into the intestines. If this occurs, a plastic or self-expanding metal (SEM) stent is placed to restore bile drainage. United States (US) Food and Drug Administration (FDA)-approved plastic biliary stents are less expensive than SEMs but have limited patency and can occlude bile flow if placed spanning a duct juncture. Recently, we investigated the effects of variations to post-processing and autoclaving on a commercially available stereolithography (SLA) resin in an effort to produce a suitable material for use in a biliary stent, an FDA Class II medical device. We tested six variations from the manufacturer's recommended post-processing and found that tripling the isopropanol (IPA) wash time to 60 min and reducing the time and temperature of the UV cure to 10 min at 40 °C, followed by a 30 min gravity autoclave cycle, yielded a polymer that was flexible and non-cytotoxic. In turn, we designed and fabricated customizable, SLA 3D-printed polymeric biliary stents that permit bile flow at a duct juncture and can be deployed via catheter. Next, we generated an in silico stent 3-point bend test to predict displacements and peak stresses in the stent designs. We confirmed our simulation accuracy with experimental data from 3-point bend tests on SLA 3D-printed stents. Unfortunately, our 3-point bend test simulation indicates that, when bent to the degree needed for placement via catheter (~30°), the peak stress the stents are predicted to experience would exceed the yield stress of the polymer. Thus, the risk of permanent deformation or damage during placement via catheter to a stent printed and post-processed as we have described would be significant. Moving forward, we will test alternative resins and post-processing parameters that have increased elasticity but would still be compatible with use in a Class II medical device.

Indexed as

3D printingbiliary stentCFDFEApost-processingSLAsterilization