ArticleHeart rhythm2025
Appropriate and inappropriate ICD shocks in patients with LVADs: Prevalence, associated factors, and etiologies.
Article in Heart rhythm, 2025. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 2 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
2 citing papers in PubMed.
- Impact of cardiac resynchronization therapy in patients with left ventricular assist devices: A systematic review and meta-analysis.JHLT open · 2026Article
- Management of arrhythmias in left ventricular assist device recipients.Heart failure reviews · 2025Review
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
20 authors.
Funding
Abstract
backgroundImplantable cardioverter-defibrillator (ICD) shocks are a common complication after left ventricular assist device (LVAD) implantation; however, data on their frequency and causes are limited.
objectiveThe purpose of this study was to define the incidence, programming, patient characteristics, and factors associated with appropriate and inappropriate ICD shocks in persons with LVADs.
methodsWe performed a retrospective review at Duke University Hospital of all LVAD recipients implanted between January 1, 2013, to June 30, 2019, with a preexisting ICD. ICD shocks were adjudicated by the treating physician and a second reviewer for the purpose of this study.
resultsAmong 421 patients with an ICD in situ undergoing LVAD implant, 147 (33.9%) had at least 1 shock after LVAD implantation. Among 134 patients with complete device history, a total of 330 shock episodes occurred: 255 (77.3%) appropriate and 75 (22.7%) inappropriate. Etiologies for inappropriate shocks included supraventricular tachycardia (n = 66 [20.0%]), physiological oversensing (n = 1 [0.3%]), and nonphysiological oversensing (n = 8 [2.4%]) including LVAD electromagnetic interference (n = 1 [0.3%]). ICD programming with shorter detection delay (P <.001) and absence of antitachycardia pacing programming (P = .001) in high-rate zones was seen more commonly in inappropriate shock than appropriate shock.
conclusionsThe rate of inappropriate shocks in LVAD recipients is very high and most often is due to supraventricular arrhythmias. LVAD electromagnetic interference is a rare cause of ICD shock. Implementation of current consensus American Heart Association recommendations for LVAD programming with long detection delays and high rate cutoffs may help prevent inappropriate ICD shocks.