Performance evaluation of the 3D-ring cadmium-zinc-telluride (CZT) StarGuide system according to the NEMA NU 1-2018 standard.

Alessandra Zorz, Marco Andrea Rossato, Paolo Turco, Luca Maria Colombo Gomez, Andrea Bettinelli, Francesca De Monte, Marta Paiusco, Pietro Zucchetta, Diego Cecchin

Abstract read
In one paragraph

Article in EJNMMI physics, 2024. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 15 papers.

0numbers the graph read from it
0cells of the map it votes in
15citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

15 citing papers in PubMed.

  1. Optimisation of Q.Clear (BSREM) Reconstruction Parameters forBioengineering (Basel, Switzerland) · 2026
    Article
  2. QuantitativeEJNMMI physics · 2026
    Article
  3. Article
  4. Article
  5. Article
  6. Review
  7. Quantitative SPECT/CT: Technique and Clinical Applications.Recent results in cancer research. Fortschritte der Krebsforschung. Progres dans les recherches sur le cancer · 2026
    Review
  8. An European consensus on [EJNMMI physics · 2025
    Article
  9. Article
  10. Article
  11. Article
  12. Article
  13. Article
  14. Review
  15. Impact of Collimator Material on Spatial Resolution and Sensitivity in Semiconductor-Based Imaging Systems: A Monte Carlo Evaluation.Indian journal of nuclear medicine : IJNM : the official journal of the Society of Nuclear Medicine, India
    Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

9 authors.

Alessandra ZorzMedical Physics Department, Veneto Institute of Oncology IOV - IRCCS, Padua, Italy. alessandra.zorz@iov.veneto.it.ORCID http://orcid.org/0000-0001-5650-9832
Marco Andrea RossatoMedical Physics Department, Veneto Institute of Oncology IOV - IRCCS, Padua, Italy.
Paolo TurcoUnit of Nuclear Medicine, University Hospital of Padova, Padua, Italy.
Luca Maria Colombo GomezMedical Physics Department, Veneto Institute of Oncology IOV - IRCCS, Padua, Italy.
Andrea BettinelliMedical Physics Department, Veneto Institute of Oncology IOV - IRCCS, Padua, Italy.
Francesca De MonteMedical Physics Department, Veneto Institute of Oncology IOV - IRCCS, Padua, Italy.
Marta PaiuscoMedical Physics Department, Veneto Institute of Oncology IOV - IRCCS, Padua, Italy.
Pietro ZucchettaUnit of Nuclear Medicine, University Hospital of Padova, Padua, Italy.
Diego CecchinUnit of Nuclear Medicine, Department of Medicine (DIMED), University Hospital of Padova, Padua, Italy.

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

backgroundThe application of semi-conductor detectors such as cadmium-zinc-telluride (CZT) in nuclear medicine improves extrinsic energy resolution and count sensitivity due to the direct conversion of gamma photons into electric signals. A 3D-ring pixelated CZT system named StarGuide was recently developed and implemented by GE HealthCare for SPECT acquisition. The system consists of 12 detector columns with seven modules of 16 × 16 CZT pixelated crystals, each with an integrated parallel-hole tungsten collimator. The axial coverage is 27.5 cm. The detector thickness is 7.25 mm, which allows acquisitions in the energy range [40-279] keV. Since there is currently no performance characterization specific to 3D-ring CZT SPECT systems, the National Electrical Manufacturers Association (NEMA) NU 1-2018 clinical standard can be tailored to these cameras. The aim of this study was to evaluate the performance of the SPECT/CT StarGuide system according to the NEMA NU 1-2018 clinical standard specifically adapted to characterize the new 3D-ring CZT.

resultsDue to the integrated collimator, the system geometry and the pixelated nature of the detector, some NEMA tests have been adapted to the features of the system. The extrinsic measured energy resolution was about 5-6% for the tested isotopes (

conclusionsThe performance tests of the StarGuide can be performed according to the NEMA NU 1-2018 standard with some adaptations. The system has shown promising results, particularly in terms of energy resolution, spatial resolution and volumetric sensitivity, potentially leading to higher quality clinical images.

Indexed as

CZTNEMAQuality controlSPECT/CT