Evidence map›Paper›PMID 39042435›Full record

ArticleJournal of applied clinical medical physics2024

Off-iso Winston-Lutz test on seven linear accelerators.

Junfang Gao, David Anand

Abstract read
In one paragraph

Article in Journal of applied clinical medical physics, 2024. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 7 papers.

0numbers the graph read from it
0cells of the map it votes in
7citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

7 citing papers in PubMed.

  1. Article
  2. Article
  3. Article
  4. Five-year evaluation of linear accelerator-based SRS platform isocentricity.Journal of applied clinical medical physics · 2025
    Article
  5. Article
  6. Off-iso Winston-Lutz test on seven linear accelerators.Journal of applied clinical medical physics · 2024
    Article
  7. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

2 authors.

Junfang GaoRadiotherapy Clinics of Georgia, Decatur, Georgia, USA.ORCID https://orcid.org/0000-0002-9056-9983
David AnandRadiation oncology department, Texas Oncology, Houston, Texas, USA.

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

purposeThe aim of this study is to find optimal gantry, collimator, and couch angles for performing single isocenter, multiple target stereotactic radiosurgery (SIMT-SRS). Nineteen angle sets were tested across seven linear accelerators for radiation-isocenter coincidence and off-isocenter coincidence. The off-isocenter Winston-Lutz test was performed to evaluate the accuracy of isocenter alignment for each angle set, and optimal angle sets as well as maximum off-isocenter distance to target for each angle set was determined. The influence of simulated patient weight on off-iso Winston-Lutz test accuracy was also inspected.

methodThe SNC MultiMet-WL phantom and MultiMet-WL QA Software v2.1 were used for the direct measurement and analysis of the off-iso Winston-Lutz test (also referred to as Winston-Lutz-Gao test). A two-step method was developed to ensure precise initial placement of the target. Nineteen beams were delivered at 6X energy and 2 × 2 cm field size to each of six targets on the MultiMet Cube with couch kicks at five cardinal angles (90°, 45°, 0°, 315°, and 270°). To reduce imaging uncertainty, only EPID was used in target alignment and test image acquisition. A total of 200 Ibs (90.7 kg) of weight was also used to mimic patient weight. All tests were performed on both the free table and the weighted table.

resultsFor two new TrueBeam machines, the maximum offset was within the 1 mm tolerance when the off-iso distance is less than 7 cm. Two older VitalBeam machines exhibited unfavorable gantry, couch, and collimator (GCC) angle sets: Linac No. 3 at (0,90,0), (0,270,0) and Linac No. 4 at (0,45,45) and (0,90,0). The C-Series Linacs failed in the majority of GCC angle sets, with Linac No. 5 exhibiting a maximum offset of 1.53 mm. Four of seven machines show a clear trend that offset increases with off-isocenter distance. Additionally, the IGRT table was less susceptible to the addition of simulated patient weight than the ExactCouch.

conclusionAmong the seven linear accelerators addressed, newer model machines such as the Varian TrueBeam were more precise than older models, especially in comparison to the C-Series Linacs. The newer machines are more suitable for delivering SIMT-SRS procedures in all GCC angle sets, and the results indicate that newer TrueBeams are capable of performing SIMT-SRS procedures at all angle sets for targets of off-iso distances up to 7 cm. The trend that offset between the target center and radiation field center increases with off-iso distance, however, does not always hold true across machines. This may be comprised by the EPID's severe off-axis horn effect. Lastly, the IGRT couch was less susceptible to patient weight compared to ExactCouch in the off-isocenter Winston-Lutz test.

Indexed as

Particle AcceleratorsPhantoms, ImagingRadiosurgeryRadiotherapy DosageRadiotherapy, Intensity-ModulatedRadiotherapy Planning, Computer-AssistedHumansNeoplasmsSoftwaremultiple‐mets SRSoff‐iso Winston‐Lutz testSIMT‐SRSWinston‐Lutz‐Gao test

Identifiers

PMID39042435
PMCPMC11466459

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.