ArticleBMC ophthalmology2024
Clinical outcomes of conjunctivochalasis treatment with a new ophthalmic radiofrequency device.
Article in BMC ophthalmology, 2024. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 2 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
2 citing papers in PubMed.
- Current and Emerging Approaches in the Management of Severe Ocular Surface Disease.Medicina (Kaunas, Lithuania) · 2025Review
- Efficacy of a novel surgical approach: sutureless correction of conjunctivochalasis using new conjunctival forceps combined with high-frequency electrocautery.Frontiers in ophthalmology · 2025Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
4 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
purposeTo investigate the safety and efficacy of a new micro-controlled radiofrequency device for treatment of conjunctivochalasis (Cch).
methodsData of 127 patients (230 eyes) who underwent ophthalmic radiofrequency treatment for Cch from January 2020 to June 2023 were analyzed retrospectively. Cch coagulation was performed with a radiofrequency electrode tip (OcuRF®, Ilooda, Korea) and a high-frequency radio-wave electric unit (0.6 ~ 0.8 watts, 2 MHz, Acutron™, Ilooda, Korea). Pre- and postoperative Cch grading, slit-lamp photography, tear film break-up time (TBUT), and bulbar conjunctival hyperemia using Keratograph 5 M (Oculus, Wetzlar, Germany) were evaluated. Cch grade 0 or 1 after surgery was regarded as 'success'. Complications, recurrence, and additional treatment rates were analyzed.
resultsIn 227 (98.7%) eyes, the radiofrequency treatment led to marked improvement of Cch, with 224 (97.4%) eyes achieving grade 0 or 1 at 2 months postoperatively. Eight eyes (3.5%) received additional treatment. TBUT improved from 3.17 ± 0.82 s to 5.28 ± 1.10 s after surgery (P < 0.001). The total bulbar conjunctival hyperemia value showed an improvement from 1.7 ± 0.6 to 1.4 ± 0.6 postoperatively (P < 0.05). No serious complications were observed.
conclusionThe novel ophthalmic radiofrequency device led to a marked improvement of Cch with no serious adverse events during the entire follow-up period. Our results suggest that the radiofrequency device presents a safe and efficacious treatment option for Cch.