ArticleCardiovascular engineering and technology2024
Early-stage Development of the CoRISMA Mechanical Circulatory Support (CMCS) System for Heart Failure Therapy.
Article in Cardiovascular engineering and technology, 2024. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 4 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
4 citing papers in PubMed.
- Feasibility Testing of the 2nd Generation Inspired Therapeutics NeoMate System for Paracorporeal Support of Neonates and Infants in Heart Failure.Artificial organs · 2026Article
- Flow benefits of pediatric arterial cannulae versus introducer sheaths as distal perfusion catheters in venoarterial extracorporeal membrane oxygenation.JTCVS open · 2025Article
- Hemodynamic considerations of ipsilateral versus contralateral cannulation with venoarterial extracorporeal membrane oxygenation.JTCVS open · 2025Article
- Feasibility Testing of the Bionet Sonar Ultrasound Transcutaneous Energy Transmission (UTET) System for Wireless Power and Communication of a LVAD.Cardiovascular engineering and technology · 2024Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
10 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
purposeCoRISMA MCS Systems Inc (Hamden CT) is developing an innovative mechanical circulatory support system (CMCS) as a durable therapeutic option for heart failure (HF) patients. The CMCS system is comprised of an axial flow pump, non-contacting hydrodynamic bearings, and integrated DC motor designed to be fully implantable in a left atrial (LA) to aortic (Ao) configuration; this unloading strategy may be particularly beneficial for HF patients with preserved ejection fraction (HFpEF). The small (5.5 cm
methodsComputational fluid dynamics (CFD) modeling was performed to evaluate flow and shear in two impeller (3 blades, 0.5 mm thickness, 8.9 mm diameter, 0.15 mm gap, polished titanium) and diffusor (5 blades, polished titanium) candidate designs. Test apparatuses were custom built to expedite development of the impeller/diffuser designs and iteratively refine the CFD models. Two candidate impeller/diffusor designs were fabricated and tested in each of the two test apparatuses (n = 4 impeller/diffuser + test fixture configurations) in static mock flow loops (hydrodynamic H-Q curves, 3.5 cP glycerol solution at 37 °C), and in dynamic mock flow loops (hemodynamics, 3.5 cP glycerol solution at 37 °C) tuned to HF conditions (mean aortic pressure 50 mmHg, central venous pressure 15 mmHg, aortic flow 3.0 L/min, and heart rate 80 bpm).
resultsCFD predicted flows of 4.56 L/min and 4.82 L/min at 100 mmHg for impellers/diffusers 1 and 2, respectively. Impeller 2 required less torque to generate a 6% increase in fluidic flow, and the diffuser had a larger area of high pressure, indicative of lower friction, which likely contributed to the increased efficiency. Experimental testing for all four configurations in the static and dynamic mock loops met performance metrics as evidenced by generating 4.0-4.5 L/min flow against 70-76 mmHg pressure at 25,000 rpm and restoring hemodynamics in the dynamic mock flow loop (MAP = 80 mmHg, CVP = 0 mmHg, total flow = 5.5 L/min) from baseline simulated HF test conditions.
conclusionThese results demonstrate proof-of-concept of the early engineering design and performance of the prototype CoRISMA pumps. Engineering specifications, challenges observed, and proposed solutions for the next design iteration were identified for the continued development of an effective, reliable, and safe LA-to-Ao CMCS system for HF patients. Current design plans are underway for incorporating a wireless energy transfer system for communication and power, eliminating the need for and complications associated with an external driveline, to achieve a fully-implantable system.