Continuous non-invasive vs. invasive arterial blood pressure monitoring during neuroradiological procedure: a comparative, prospective, monocentric, observational study.
Article in Perioperative medicine (London, England), 2024. The graph could read no effect estimate from its abstract, so it casts no vote on the map. It is linked to trial NCT05283824 (Comparison of 3 Techniques for Blood Pressure Measurements During Interventional Neuroradiology Procedures), which is not on this map. Cited by 1 paper.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Comparison of 3 Techniques for Blood Pressure Measurements During Interventional Neuroradiology Procedures
Who cites it
1 citing paper in PubMed.
- Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
8 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
backgroundIn the perioperative setting, the most accurate way to continuously measure arterial blood pressure (ABP) is using an arterial catheter. Surrogate methods such as finger cuff have been developed to allow non-invasive measurements and are increasingly used, but need further evaluation. The aim of this study is to evaluate the accuracy and clinical concordance between two devices for the measurement of ABP during neuroradiological procedure.
methodsThis is a prospective, monocentric, observational study. All consecutive patients undergoing a neuroradiological procedure were eligible. Patients who needed arterial catheter for blood pressure measurement were included. During neuroradiological procedure, ABP (systolic, mean and diatolic blood pressure) was measured with two different technologies: radial artery catheter and Nexfin. Bland-Altman and error grid analyses were performed to evaluate the accuracy and clinical concordance between devices.
resultsFrom March 2022 to November 2022, we included 50 patients, mostly ASA 3 (60%) and required a cerebral embolization (94%) under general anaesthesia (96%). Error grid analysis showed that 99% of non-invasive ABP measures obtained with the Nexfin were located in the risk zone A or B. However, 65.7% of hypertension events and 41% of hypotensive events were respectively not detected by Nexfin. Compared to the artery catheter, a significant relationship was found for SAP (r
conclusionsNexfin is not strictly interchangeable with artery catheter for ABP measuring. Further studies are needed to define its clinical use during neuroradiological procedure.
trial registrationClinicaltrials.gov, registration number: NCT05283824.