Evaluation of the measurement accuracy and uncertainty of a solid-state detector under diagnostic x-ray beam conditions.
Article in Journal of applied clinical medical physics, 2024. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 4 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
4 citing papers in PubMed.
- Basic characteristics of do-it-yourself dosimeter in general radiography system.Radiological physics and technology · 2026Article
- Solid-state dosimeters have rendered ionization chambers obsolete for routine diagnostic X-ray quality control.Journal of applied clinical medical physics · 2026Article
- X-ray multimeter performance under calibration laboratory conditions.Medical physics · 2025Article
- Evaluation of the measurement accuracy and uncertainty of a solid-state detector under diagnostic x-ray beam conditions.Journal of applied clinical medical physics · 2024Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
4 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
objectiveAn accurate measurement of x-ray beams is expected to reduce the uncertainties associated with estimating radiation risk to patients in clinical settings. To perform assessment tasks based on the readings of a solid-state detector (SSD) using semiconductor technology, the characteristics of the detector should be elucidated. In this study, we evaluated the measurement accuracy of a new SSD under diagnostic x-ray beam conditions in terms of air kerma, tube voltage, and half-value layer (HVL). The performance of the SSD was then compared with those of reference instruments.
methodsThe tube voltage was varied within the range of 50-120 kV in steps of 10 kV and the thickness and materials of additional filters were concurrently changed (several combinations were tested). In addition, the dose rate and energy dependence of the SSD were also investigated. These effects were analyzed based on statistical significance tests. Furthermore, the expanded uncertainties in the series of measurements were meticulously calculated.
resultsThe results showed average relative differences of -3.26 ± 1.33%, 0.44 ± 1.01%, and -2.60 ± 3.31% for air kerma, tube voltage, and HVL, respectively. Furthermore, air kerma did not exhibit any dependence on dose rate and energy, in contrast to tube voltage and HVL measurements.
conclusionThe measurement values of the SSD fall within the acceptable range of uncertainty, highlighting its measurement accuracy and reliability. Furthermore, based on the characteristics elucidated by this study, valuable insights are provided concerning the assurance of appropriate measurement values in clinical settings.