Minimal effect of scanning parameters on ultrasound shear wave elastography variability in tendons.
Article in Knee surgery, sports traumatology, arthroscopy : official journal of the ESSKA, 2024. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 2 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
2 citing papers in PubMed.
- Ultrasound shear wave elastography - a reliable protocol for the assessment of the stiffness of the supraspinatus tendon.Shoulder & elbow · 2025Article
- Effect of joint angle positioning on shearwave speed and variability with ultrasound shearwave elastography in asymptomatic Achilles and patellar tendons.Journal of biomechanics · 2024Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
10 authors.
Funding
Abstract
purposeUltrasound shear wave elastography has potential use in assessing tendon tissue; however, reducing measurement variability remains challenging. The primary purpose of this study was to identify the amount of variability accounted for by ultrasound parameter (frequency, harmonics and CrossXBeam) settings on shear wave speed at two testing sites.
methodsShear wave elastography images of the Achilles tendon were obtained from individuals with healthy tendons (n = 28) at two testing sites with standardised image acquisition/postprocessing protocols. Images were acquired at a range of frequencies (7-15 MHz) with CrossXBeam (a filtering technique) and harmonics settings toggled on and off. Variance decomposition analysis was performed to identify the amount of variability in shear wave speed accounted for by scan acquisition settings and testing sites.
resultsShear wave speed variance was primarily attributed to participants (56.87% of variance; residual error: 35%). All scanning parameters, testing site and interaction terms each accounted for less than 2.5% of the variance. A statistically significant, negative relationship was observed between shear wave speed and image quality (p = 0.001) suggesting poor image quality yields higher shear wave speed estimates.
conclusionThe findings of this study suggest that natural variation in Achilles tendon mechanics between individuals without tendon pathology accounts for most of the shear wave speed variability. Optimising image quality, which may be observed in higher frequencies, should be considered to improve shear wave speed estimation. Clinically, this study highlights the need to take multiple images, maintain consistent ultrasound settings when tracking patient progress over time and use caution when comparing raw values from tendon scans performed in different clinics with shear wave elastography. LEVEL OF EVIDENCE: Level III.