ArticleBlood pressure monitoring2025

Validation of the DBP-8278B Wrist-type Fully Automatic Blood Pressure Monitor in the general population according to the ISO 81060-2:2018/AMD 1:2020 protocol.

Qi Zhou

Abstract readValidation Study
In one paragraph

Article in Blood pressure monitoring, 2025. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 1 paper.

0numbers the graph read from it
0cells of the map it votes in
1citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

1 citing paper in PubMed.

  1. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

1 author.

Qi ZhouCardiovascular Medicine Department, Hebei Petro China Central Hospital, Langfang, Hebei Province, China.

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

The aim of this study was to evaluate the accuracy of the DBP-8278B Wrist-type Fully Automatic Blood Pressure Monitor in the general population according to the AAMI/ESH/ISO Universal Standard (ISO 81060-2 : 2018/AMD 1 : 2020). The test monitor was used to measure SBP and DBP and pulse rate in the wrists using the oscillometric method. According to this protocol using the same-arm sequential blood pressure measurement method, subjects were recruited to fulfill the age, sex, blood pressure, and cuff distribution criteria of the Universal Standard. A single cuff for wrist sizes of 13.5-21.5 cm was used. A total of 85 subjects were included in this study. According to Criterion 1, the mean difference in SBP between the test device and the reference device was 0.89 mmHg, with a SD of 7.76 mmHg. The mean difference of DBP was -1.44 mmHg, with a SD of 6.23 mmHg. The mean difference between SBP and DBP was less than 5.0 mmHg, and the SD was less than 8.0 mmHg, which met the requirements. According to Criterion 2, the SD of the mean differences of SBP between the test device and the reference device per subject was 5.65 mmHg, which was less than 6.88 mmHg and met the requirements. The SD of the mean differences of DBP between the test device and the reference device per subject was 5.61 mmHg, which was less than 6.78 mmHg and met the requirements. The DBP-8278B Wrist-type Fully Automatic Blood Pressure Monitor met the requirements of the AAMI/ESH/ISO Universal Standard (ISO 81060-2 : 2018/AMD 1 : 2020). It can be recommended for home and clinical use.

Indexed as

Blood Pressure MonitorsWristAdultAgedBlood PressureBlood Pressure DeterminationFemaleHumansMaleMiddle Aged

Identifiers

PMID39008688
PMCPMC11654442

What OpenQuestion holds

Textmetadata
LicenceCC BY-NC-ND
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.