ArticleBlood pressure monitoring2025
Validation of the DBP-8278B Wrist-type Fully Automatic Blood Pressure Monitor in the general population according to the ISO 81060-2:2018/AMD 1:2020 protocol.
Article in Blood pressure monitoring, 2025. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 1 paper.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
1 citing paper in PubMed.
- Credence of device acceptability: a statistical method for comparing blood pressure measurement device accuracy across studies.Journal of human hypertension · 2025Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
1 author.
Funding
No grant is acknowledged in the PubMed record.
Abstract
The aim of this study was to evaluate the accuracy of the DBP-8278B Wrist-type Fully Automatic Blood Pressure Monitor in the general population according to the AAMI/ESH/ISO Universal Standard (ISO 81060-2 : 2018/AMD 1 : 2020). The test monitor was used to measure SBP and DBP and pulse rate in the wrists using the oscillometric method. According to this protocol using the same-arm sequential blood pressure measurement method, subjects were recruited to fulfill the age, sex, blood pressure, and cuff distribution criteria of the Universal Standard. A single cuff for wrist sizes of 13.5-21.5 cm was used. A total of 85 subjects were included in this study. According to Criterion 1, the mean difference in SBP between the test device and the reference device was 0.89 mmHg, with a SD of 7.76 mmHg. The mean difference of DBP was -1.44 mmHg, with a SD of 6.23 mmHg. The mean difference between SBP and DBP was less than 5.0 mmHg, and the SD was less than 8.0 mmHg, which met the requirements. According to Criterion 2, the SD of the mean differences of SBP between the test device and the reference device per subject was 5.65 mmHg, which was less than 6.88 mmHg and met the requirements. The SD of the mean differences of DBP between the test device and the reference device per subject was 5.61 mmHg, which was less than 6.78 mmHg and met the requirements. The DBP-8278B Wrist-type Fully Automatic Blood Pressure Monitor met the requirements of the AAMI/ESH/ISO Universal Standard (ISO 81060-2 : 2018/AMD 1 : 2020). It can be recommended for home and clinical use.