Observational studyBMC anesthesiology2024
Accuracy of oscillometric noninvasive blood pressure at the ankle in the lateral position during general anesthesia.
Observational study in BMC anesthesiology, 2024. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 1 paper.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
1 citing paper in PubMed.
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
7 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
backgroundThis study aimed to evaluate the accuracy of ankle blood pressure measurements in relation to invasive blood pressure in the lateral position.
methodsThis prospective observational study included adult patients scheduled for elective non-cardiac surgery under general anesthesia in the lateral position. Paired radial artery invasive and ankle noninvasive blood pressure readings were recorded in the lateral position using GE Carescape B650 monitor. The primary outcome was the ability of ankle mean arterial pressure (MAP) to detect hypotension (MAP < 70 mmHg) using area under the receiver operating characteristic curve (AUC) analysis. The secondary outcomes were the ability of ankle systolic blood pressure (SBP) to detect hypertension (SBP > 140 mmHg) as well as bias (invasive measurement - noninvasive measurement), and agreement between the two methods using the Bland-Altman analysis.
resultsWe analyzed 415 paired readings from 30 patients. The AUC (95% confidence interval [CI]) of ankle MAP for detecting hypotension was 0.88 (0.83-0.93). An ankle MAP of ≤ 86 mmHg had negative and positive predictive values (95% CI) of 99 (97-100)% and 21 (15-29)%, respectively, for detecting hypotension. The AUC (95% CI) of ankle SBP to detect hypertension was 0.83 (0.79-0.86) with negative and positive predictive values (95% CI) of 95 (92-97)% and 36 (26-46)%, respectively, at a cutoff value of > 144 mmHg. The mean bias between the two methods was - 12 ± 17, 3 ± 12, and - 1 ± 11 mmHg for the SBP, diastolic blood pressure, and MAP, respectively.
conclusionIn patients under general anesthesia in the lateral position, ankle blood pressure measurements are not interchangeable with the corresponding invasive measurements. However, an ankle MAP > 86 mmHg can exclude hypotension with 99% accuracy, and an ankle SBP < 144 mmHg can exclude hypertension with 95% accuracy.