ArticleInternational ophthalmology2024
Accuracy of a novel method for IOP measurement without applying pressure on the cornea.
Article in International ophthalmology, 2024. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Not yet cited in PubMed.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
0 citing papers in PubMed.
No citing paper in PubMed yet.
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
5 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
purposeTo assess the accuracy of a prototype novel instrument for intra ocular pressure (IOP) measurements not involving corneal pressure application.
designProspective case control study.
methodsAn institutional study including 16 healthy volunteers without ocular pathology. IOP in both eyes of the participants was measured four times in different body positions with the novel prototype and reference instrument (Goldmann applanation tonometer (GAT) or iCare (iCare Finland OY, Vantaa, Finland)). IOP results were compared between the prototype and the reference instruments in 116 pairs of measurement.
resultsOverall no statistically significant difference was found between the presented prototype and the reference instrument. Stratifying measurements by instrument used revealed no significant difference for GAT and statistical significant (yet clinically insignificant) difference for iCare.
conclusionsThe presented prototype demonstrates good clinical agreement of IOP measuring results with reference instruments Further large-scale studies assessing this instrument in glaucoma patients are warranted.