ArticlePloS one2024
Deformation investigation of electromagnetic diaphragm pump rubber diaphragm.
Article in PloS one, 2024. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Not yet cited in PubMed.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
0 citing papers in PubMed.
No citing paper in PubMed yet.
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
4 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
A diaphragm pump is a type of volumetric pump that has excellent sealing performance. An electromagnetic diaphragm pump is a kind of widely adopted diaphragm pump that has a simple structure, low power loss, and high cost performance. However, the calculation method of deformation for the electromagnetic diaphragm pump rubber diaphragm is presently lacking. Herein, a calculating method of deformation for the electromagnetic diaphragm pump rubber diaphragm is proposed. By establishing and analyzing a deformation model of the electromagnetic diaphragm pump rubber diaphragm, a theoretical relationship between the deformation of the electromagnetic diaphragm pump rubber diaphragm, the size of the electromagnetic diaphragm pump rubber diaphragm and the pressure of fluid is determined. The experimental results indicate that the biggest difference between the tested axial deformation and the calculated axial deformation of the electromagnetic diaphragm pump rubber diaphragm is 0.04 mm and the calculation results show agreement with the experimental results.
Indexed as
Identifiers
What OpenQuestion holds
Registered trials
Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.