Retrieval analysis of PEEK rods pedicle screw system: three cases analysis.

Xiaoduo Xu, Lei Wang, Jingming Wang, Xiuchun Yu, Weimin Huang

Abstract readCase Reports
In one paragraph

Article in BMC musculoskeletal disorders, 2024. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 2 papers.

0numbers the graph read from it
0cells of the map it votes in
2citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

2 citing papers in PubMed.

  1. Review
  2. Review
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

5 authors.

Xiaoduo XuOrthopedic Department, 960 Hospital of People's Liberation Army, NO.25 Shifan Road, Jinan, Shandong, 250031, China.
Lei WangOrthopedic Department, 960 Hospital of People's Liberation Army, NO.25 Shifan Road, Jinan, Shandong, 250031, China.
Jingming WangOrthopedic Department, 960 Hospital of People's Liberation Army, NO.25 Shifan Road, Jinan, Shandong, 250031, China.
Xiuchun YuOrthopedic Department, 960 Hospital of People's Liberation Army, NO.25 Shifan Road, Jinan, Shandong, 250031, China. 13969132190@163.com.
Weimin HuangOrthopedic Department, 960 Hospital of People's Liberation Army, NO.25 Shifan Road, Jinan, Shandong, 250031, China. ever_23@163.com.

Funding

Jinan Clinical Medical Science and Technology Innovation Plan 202134006
6 · The paper itself

Abstract

purposeTo analyze the characteristics of PEEK rods retrieved in vivo, specifically their wear and deformation, biodegradability, histocompatibility, and mechanical properties.

methodSix PEEK rods were retrieved from revision surgeries along with periprosthetic tissue. The retrieved PEEK rods were evaluated for surface damage and internal changes using Micro-CT, while light and electron microscopy were utilized to determine any histological changes in periprosthetic tissues. Patient history was gathered from medical records. Two intact and retrieved PEEK rods were used for fatigue testing analysis by sinusoidal load to the spinal construct.

resultsAll implants showed evidence of plastic deformation around the screw-rod interface, while the inner structure of PEEK rods appeared unchanged with no visible voids or cracks. Examining images captured through light and electron microscopy indicated that phagocytosis of macrophages around PEEK rods was less severe in comparison to the screw-rod interface. The results of an energy spectrum analysis suggested that the distribution of tissue elements around PEEK rods did not differ significantly from normal tissue. During fatigue testing, it was found that the retrieved PEEK rods cracked after 1.36 million tests, whereas the intact PEEK rods completed 5 million fatigue tests without any failure.

conclusionPEEK rods demonstrate satisfactory biocompatibility, corrosion resistance, chemical stability, and mechanical properties. Nevertheless, it is observed that the indentation at the junction between the nut and the rod exhibits relatively weak strength, making it susceptible to breakage. As a precautionary measure, it is recommended to secure the nut with a counter wrench, applying the preset torque to prevent overtightening.

Indexed as

BenzophenonesKetonesPedicle ScrewsPolymersAgedBiocompatible MaterialsDevice RemovalFemaleHumansMaleMaterials TestingMiddle AgedPolyethylene GlycolsProsthesis FailureReoperationBenzophenonesBiocompatible MaterialsKetonespolyetheretherketonePolyethylene GlycolsPolymersFatigue testingHistological responsePEEK rodsRetrieval analysisRevision surgery

Identifiers

PMID38909212
PMCPMC11193296

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.