ArticleIEEE transactions on bio-medical engineering2024
A Clinically Feasible Electrode Array for 3D Intraoperative Electrical Impedance Tomography-Based Surgical Margin Assessment in Robot-Assisted Radical Prostatectomy.
Article in IEEE transactions on bio-medical engineering, 2024. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 3 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
3 citing papers in PubMed.
- In Vivo Classification of Oral Lesions Using Electrical Impedance Spectroscopy.IEEE transactions on bio-medical engineering · 2026Article
- Comparison of Complex Open Domain Electrical Impedance Tomography Methods.IEEE transactions on bio-medical engineering · 2025Article
- Induction of subject-ventilator asynchrony by variation of respiratory parameters in a lung injury model in pigs.Respiratory research · 2024Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
5 authors.
Funding
Abstract
objectiveA novel small form factor circular electrode array was designed specifically for electrical impedance tomography (EIT) based assessment of surgical margins during robot assisted radical prostatectomy (RARP).
methodsThe electrode array consists of 33 gold-plated electrodes arranged within a 9.5 mm diameter circular footprint on the end of a surgical probe that can be introduced through a standard 12 mm laparoscopic port used during RARP. The electrode array contains 8 larger, low-contact impedance outer electrodes dedicated for current drive and an internal grid of 25 smaller electrodes for simultaneous voltage measurement. Separating electrode geometry by function is designed to improve current delivery, speed, and resolution while reducing hardware requirements.
resultsSimulations demonstrated that 1 mm diameter hemispherical prostate cancer inclusions could be localized within regions of adipose and benign prostate tissue; 1.5 mm diameter inclusions were required for localization within muscle tissue. A 2.38 mm diameter aluminum rod in 0.2 S/m saline could be localized throughout the imaging domain with a position error of less than 2.5 mm for depths from the electrode array surface of up to 1.7 mm. Ex vivo tissue experiments with a bovine model demonstrate visual congruence of muscle and adipose tissue locations between the sample and reconstructed images.
conclusionSimulation and experimental results indicate good detection and location of inclusions. SIGNIFICANCE: These results suggest the proposed electrode array design can provide sufficient accuracy in the detection and localization of prostate cancer against clinically relevant background tissues for use during RARP.