ArticleJournal of applied clinical medical physics2024
Evaluating intra-fractional tumor motion in lung stereotactic radiotherapy with deep inspiration breath-hold.
Article in Journal of applied clinical medical physics, 2024. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 8 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
8 citing papers in PubMed.
- Impact of surface-guided prepositioning and respiratory coaching on the target localization accuracy in lung stereotactic body radiation therapy.Radiological physics and technology · 2026Article
- Mid-ventilation PTV approach for stereotactic ablative body radiotherapy in stage I non-small-cell lung carcinoma: Implementation of a single-system workflow.Technical innovations & patient support in radiation oncology · 2026Article
- Effects of breath-hold reproducibility on proton and photon lung cancer stereotactic body radiotherapy.Physics and imaging in radiation oncology · 2026Article
- Dosimetric and treatment efficiency comparison of lung SBRT using three different motion management strategies.Clinical and translational radiation oncology · 2026Article
- Quantifying interplay effects in lung cancer IMPT: a comprehensive analysis of treatment planning parameter sensitivity.BMC cancer · 2025Article
- Dosimetric evaluation of cone beam computed tomography-guided online adaptive radiotherapy in gastric mucosa-associated lymphoid tissue lymphoma.Technical innovations & patient support in radiation oncology · 2025Article
- Application of respiratory motion management technology for patients with lung cancer treated with stereotactic body radiotherapy (Review).Oncology letters · 2025Review
- Evaluating intra-fractional tumor motion in lung stereotactic radiotherapy with deep inspiration breath-hold.Journal of applied clinical medical physics · 2024Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
8 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
purposeTo evaluate the intra-fractional tumor motion in lung stereotactic body radiotherapy (SBRT) with deep inspiration breath-hold (DIBH), and to investigate the adequacy of the current planning target volume (PTV) margins.
methodsTwenty-eight lung SBRT patients with DIBH were selected in this study. Among the lesions, twenty-three were at right or left lower lobe, two at right middle lobe, and three at right or left upper lobe. Post-treatment gated cone-beam computed tomography (CBCT) was acquired to quantify the intra-fractional tumor shift at each treatment. These obtained shifts were then used to calculate the required PTV margin, which was compared with the current applied margin of 5 mm margin in anterior-posterior (AP) and right-left (RL) directions and 8 mm in superior-inferior (SI) direction. The beam delivery time was prolonged with DIBH. The actual beam delivery time with DIBH (T
resultsA total of 113 treatments were analyzed. At six treatments (5.3%), the shifts exceeded the tolerance defined by the current PTV margin. The average shifts were 0.0 ± 1.9 mm, 0.1±1.5 mm, and -0.5 ± 3.7 mm in AP, RL, and SI directions, respectively. The required PTV margins were determined to be 4.5, 3.9, and 7.4 mm in AP, RL, and SI directions, respectively. The average T
conclusionIntra-fractional tumor motion is the predominant source of treatment uncertainties in CBCT-guided lung SBRT with DIBH. The required PTV margin should be determined based on data specific to each institute, considering different techniques and populations. Our data indicate that our current applied PTV margin is adequate, and it is possible to reduce further in the RL direction. The time increase of T