Evidence map›Paper›PMID 38780791›Full record

ArticleEuropean journal of orthopaedic surgery & traumatology : orthopedie traumatologie2024

Femoral neck osteosynthesis using interfragmentary compression or position screw in synthetic bones. What are the mechanical differences?

Arthur de Góes Ribeiro, Pedro Henrique Almeida, Fabio Raia, Ralph Walter Christian, José Octávio Soares Hungria, Cláudio Santili, Marcelo Tomanik Mercadante

Abstract readComparative Study
PubMed Publisher
In one paragraph

Article in European journal of orthopaedic surgery & traumatology : orthopedie traumatologie, 2024. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Not yet cited in PubMed.

0numbers the graph read from it
0cells of the map it votes in
0citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

0 citing papers in PubMed.

No citing paper in PubMed yet.

4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

7 authors.

Arthur de Góes RibeiroOrthopaedics and Traumatology Department, Irmandade da Santa Casa de Misericórdia de São Paulo, São Paulo, Brazil. arthurgr@gmail.com.ORCID http://orcid.org/0000-0002-9563-0237
Pedro Henrique AlmeidaOrthopaedics and Traumatology Department, Irmandade da Santa Casa de Misericórdia de São Paulo, São Paulo, Brazil.ORCID http://orcid.org/0009-0009-5874-3724
Fabio RaiaMackenzie Presbiterian University, São Paulo, Brazil.ORCID http://orcid.org/0000-0002-3136-3002
Ralph Walter ChristianOrthopaedics and Traumatology Department, Irmandade da Santa Casa de Misericórdia de São Paulo, São Paulo, Brazil.ORCID http://orcid.org/0000-0003-2851-9747
José Octávio Soares HungriaOrthopaedics and Traumatology Department, Irmandade da Santa Casa de Misericórdia de São Paulo, São Paulo, Brazil.ORCID http://orcid.org/0000-0003-2215-9103
Cláudio SantiliOrthopaedics and Traumatology Department, Irmandade da Santa Casa de Misericórdia de São Paulo, São Paulo, Brazil.ORCID http://orcid.org/0000-0003-0645-5500
Marcelo Tomanik MercadanteOrthopaedics and Traumatology Department, Irmandade da Santa Casa de Misericórdia de São Paulo, São Paulo, Brazil.ORCID http://orcid.org/0000-0002-3140-7562

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

purposeWe aim to compare interfragmentary compression with the position osteosynthesis in the fixation of different femoral neck fractures (FN) in synthetic bones subjected to vertical load.

methodsForty-two synthetic femurs were subjected to neck fractures and separated into 3 groups according to the Pauwels classification: Pauwels I had 6 units (PI); Pauwels II, 24 units, with and without comminution (PII); and Pauwels III, 12 units, with and without comminution (PIII). After, they were fixed with 2 different ways: three 7, 0 mm cannulated lag screws (CSs) versus three 6, 5 mm solid fully threaded screws (SSs). Screws positioning was oriented by the Pauwels classification: inverted triangle or crossed screws. All specimens were submitted to vertical loading until failure.

resultsThe average force was 79.4 ± 22.6 Kgf. The greatest one was recorded in model 1 (135.6 Kgf), and the lowest in model 41 (39.6 Kgf). CSs and SSs had similar resistance until failure (p = 0.2). PI showed heightened resistance and PIII showed a worse response (p < 0,01). CSs had better performance in PIII (p = 0.048). Comminution and screws orientation caused no difference on peak force (p = 0.918 and p = 0.340, respectively).

conclusionsIn synthetic bones, the resistance of a femoral neck fracture osteosynthesis using a 7, 0 mm cannulated lag screw or 6, 5 mm solid fully threaded screw are similar. There was no loss of efficiency with comminution in the femoral neck. Osteosynthesis resistance decreased with the verticalization of the fracture line and, in the more vertical ones, interfragmentary compression with CSs was more resistant than positional osteosynthesis with SSs.

Indexed as

Bone ScrewsFemoral Neck FracturesFracture Fixation, InternalBiomechanical PhenomenaFractures, ComminutedHumansMaterials TestingStress, MechanicalWeight-BearingFemoral neck fracturesFemur fractureHip fracturesInternal fixation devicesInternal fracture fixationMechanical tests

Identifiers

What OpenQuestion holds

Textmetadata
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.