Research on the correction method for radiotherapy verification plans based on displaced electronic portal imaging device.
Article in Journal of applied clinical medical physics, 2024. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 1 paper.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
1 citing paper in PubMed.
- Research on the correction method for radiotherapy verification plans based on displaced electronic portal imaging device.Journal of applied clinical medical physics · 2024Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
3 authors.
Funding
Abstract
backgroundIt has been observed that under the single isocenter conditions, the potential shifts of the electronic portal imaging devices (EPID) may be introduced when executing portal dosimetry (PD) plans for bilateral breast cancer, pleural mesothelioma, and lymphoma. These shifts are relative to the calibration positions of EPID and result in significant discrepancies in the plan verification results. PURPOSE: To explore methods including correction model and specific correction matrices to revise the data obtained from displaced EPID.
methodsTwo methods, the correction model and the specific correction matrices, were applied to correct the data. Five experiments were designed and conducted to build correction model and to validate the effectiveness of these two methods. Gamma passing rates were calculated and data profiles along X-axis and Y-axis were captured.
resultsThe gamma passing rates for the EPID-displaced IMRT validation plans after applying correction model, along with the application of specific correction matrices to VMAT and IMRT validation plans, exhibit results that are comparable to the cases with non-displaced EPID. Except for the VMAT plans applied correction model which showed larger discrepancies (0.041 ± 0.028, 0.049 ± 0.030), the other three exhibit minimal differences in discrepancy values. In all profiles, the corrected data from displaced EPID exhibit a high level of agreement with data obtained from non-displaced EPID. Good consistency is observed in actual application of the correction model and the specific correction matrices between gamma passing rates of data corrected and those of non-displaced data.
conclusionsThe proposed methods involving correction model and specific correction matrices can correct the data collected from the displaced EPID, and the gamma passing rates of the corrected data show results that are comparable to some extent with those of non-displaced data. Particularly, the results corrected by specific correction matrices closely resemble the data from non-displaced EPID.