ArticleJournal of applied clinical medical physics2024

Experimental and Monte Carlo based dosimetric investigation of a novel 3 mm radiosurgery 3 MV beam using the microSilicon detector.

Katrin Saße, Karina Albers, Peter Douglas Klassen, Neelan J Marianyagam, Georg Weidlich, M Bret Schneider, Steven Chang, John Adler, Björn Poppe, Hui Khee Looe and 1 more

Abstract read
In one paragraph

Article in Journal of applied clinical medical physics, 2024. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 2 papers.

0numbers the graph read from it
0cells of the map it votes in
2citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

2 citing papers in PubMed.

  1. Review
  2. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

11 authors.

Katrin SaßeUniversity Clinic for Medical Radiation Physics, Medical Campus Pius Hospital, Carl von Ossietzky University, Oldenburg, Germany.
Karina AlbersUniversity Clinic for Medical Radiation Physics, Medical Campus Pius Hospital, Carl von Ossietzky University, Oldenburg, Germany.
Peter Douglas KlassenBonifatius Hospital Lingen, Lingen, Germany.
Neelan J MarianyagamDepartment of Neurosurgery, Stanford University School of Medicine, Palo Alto, California, USA.
Georg WeidlichZAP Surgical Systems, San Carlos, California, USA.
M Bret SchneiderZAP Surgical Systems, San Carlos, California, USA.
Steven ChangDepartment of Neurosurgery, Stanford University School of Medicine, Palo Alto, California, USA.
John AdlerZAP Surgical Systems, San Carlos, California, USA.
Björn PoppeUniversity Clinic for Medical Radiation Physics, Medical Campus Pius Hospital, Carl von Ossietzky University, Oldenburg, Germany.
Hui Khee LooeUniversity Clinic for Medical Radiation Physics, Medical Campus Pius Hospital, Carl von Ossietzky University, Oldenburg, Germany.
Daniela EulensteinPTW-Freiburg, Freiburg, Germany.

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

backgroundThe ZAP-X system is a novel gyroscopic radiosurgical system based on a 3 MV linear accelerator and collimator cones with a diameter between 4 and 25 mm. Advances in imaging modalities to detect small and early-stage pathologies allow for an early and less invasive treatment, where a smaller collimator matching the anatomical target could provide better sparing of surrounding healthy tissue. PURPOSE: A novel 3 mm collimator cone for the ZAP-X was developed. This study aims to investigate the usability of a commercial diode detector (microSilicon) for the dosimetric characterization of this small collimator cone; and to investigate the underlying small field perturbation effects.

methodsProfile measurements in five depths as well as PDD and output ratio measurements were performed with a microSilicon detector and radiochromic EBT3 films. In addition, comprehensive Monte Carlo simulations were performed to validate the measurement observations and to quantify the perturbation effects of the microSilicon detector in these extremely small field conditions.

resultsIt is shown that the microSilicon detector enables an accurate dosimetric characterization of the 3 mm beam. The profile parameters, such as the FWHM and 20%-80% penumbra width, agree within 0.1 to 0.2 mm between film and detector measurements. The output ratios agree within the measurement uncertainty between microSilicon detector and films, whereas the comparisons of the PDD results show good agreement with the Monte Carlo simulations. The analysis of the perturbation factors of the microSilicon detector reveals a small field correction factor of approximately 3% for the 3 mm circular beam and a correction factor smaller than 1.5% for field diameters above 3 mm.

conclusionsIt could be shown that the microSilicon detector is well-suitable for the characterization of the new 3 mm circular beam of the ZAP-X system.

Indexed as

Monte Carlo MethodParticle AcceleratorsPhantoms, ImagingRadiosurgeryRadiotherapy DosageRadiotherapy Planning, Computer-AssistedComputer SimulationHumansRadiometryRadiotherapy, Intensity-ModulatedSiliconSiliconmicroSilicon detectorMonte Carlo simulationsradiosurgerysmall field dosimetryvolume effect

Identifiers

PMID38762906
PMCPMC11163500

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.