Evidence map›Paper›PMID 38762654›Full record

ArticleArchives of orthopaedic and trauma surgery2024

The effect of a collar on primary stability of standard and undersized cementless hip stems: a biomechanical study.

Manuel Kistler, Arnd Steinbrück, Florian Schmidutz, Alexander C Paulus, Boris Michael Holzapfel, Matthias Woiczinski

Abstract read
In one paragraph

Article in Archives of orthopaedic and trauma surgery, 2024. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 2 papers.

0numbers the graph read from it
0cells of the map it votes in
2citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

2 citing papers in PubMed.

  1. Article
  2. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

6 authors.

Manuel KistlerDepartment of Orthopaedics and Trauma Surgery, Musculoskeletal University Center Munich (MUM), University Hospital, LMU, Marchioninistraße 15, 81377, Munich, Germany. Manuel.Kistler@med.uni-muenchen.de.ORCID http://orcid.org/0000-0002-3602-6785
Arnd SteinbrückDepartment of Orthopaedics and Trauma Surgery, Musculoskeletal University Center Munich (MUM), University Hospital, LMU, Marchioninistraße 15, 81377, Munich, Germany.
Florian SchmidutzDepartment of Orthopaedics and Trauma Surgery, Musculoskeletal University Center Munich (MUM), University Hospital, LMU, Marchioninistraße 15, 81377, Munich, Germany.
Alexander C PaulusDepartment of Orthopaedics and Trauma Surgery, Musculoskeletal University Center Munich (MUM), University Hospital, LMU, Marchioninistraße 15, 81377, Munich, Germany.
Boris Michael HolzapfelDepartment of Orthopaedics and Trauma Surgery, Musculoskeletal University Center Munich (MUM), University Hospital, LMU, Marchioninistraße 15, 81377, Munich, Germany.
Matthias WoiczinskiDepartment of Orthopaedics and Trauma Surgery, Musculoskeletal University Center Munich (MUM), University Hospital, LMU, Marchioninistraße 15, 81377, Munich, Germany.

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

introductionAseptic loosening and periprosthetic fractures are main reasons for revision after THA. Quite different from most other stem systems, Corail cementless hip stems show better survival rates than their cemented counterpart, which can possibly be explained by the use of a collar. The study aimed to investigate primary stability with standard and undersized hip stems both collared and collarless. MATERIALS AND

methodsPrimary stability of cementless, collared and collarless, femoral stems was measured in artificial bones using both undersized and standard size. After preconditioning, 3D micromotion was measured under cyclic loading at the bone-implant interface.

resultsThe use of a collar resulted in higher micromotion within the same stem size but showed no statistically significant difference for both standard and undersized hip stems. The collared and collarless undersized stems showed no significant differences in 3D micromotion at the upper measuring positions compared to the standard stem size. Micromotion was significantly higher in the distal measuring positions, with and without collar, for the undersized stems (vs. standard collarless stem size).

conclusionThe key finding is that the collarless and collared Corail hip stems, within one stem size, showed no significant differences in primary stability. Undersized stems showed significantly higher micromotion in the distal area both with and without collar.

Indexed as

Arthroplasty, Replacement, HipHip ProsthesisProsthesis DesignBiomechanical PhenomenaHumansProsthesis FailureBiomechanicsCollarMicromotionPrimary stabilityTotal hip arthroplasty (THA)Undersizing

Identifiers

PMID38762654
PMCPMC11211124

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.