Development and performance evaluation of a solenoid valve assisted low-cost ventilator on gas exchange and respiratory mechanics in a porcine model.
Article in PloS one, 2024. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 1 paper.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
1 citing paper in PubMed.
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
4 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
introductionDuring the COVID-19 pandemic, ventilator shortages necessitated the development of new, low-cost ventilator designs. The fundamental requirements of a ventilator include precise gas delivery, rapid adjustments, durability, and user-friendliness, often achieved through solenoid valves. However, few solenoid-valve assisted low-cost ventilator (LCV) designs have been published, and gas exchange evaluation during LCV testing is lacking. This study describes the development and performance evaluation of a solenoid-valve assisted low-cost ventilator (SV-LCV) in vitro and in vivo, focusing on gas exchange and respiratory mechanics.
methodsThe SV-LCV, a fully open ventilator device, was developed with comprehensive hardware and design documentation, utilizing solenoid valves for gas delivery regulation. Lung simulator testing calibrated tidal volumes at specified inspiratory and expiratory times, followed by in vivo testing in a porcine model to compare SV-LCV performance with a conventional ventilator.
resultsThe SV-LCV closely matched the control ventilator's respiratory profile and gas exchange across all test cycles. Lung simulator testing revealed direct effects of compliance and resistance changes on peak pressures and tidal volumes, with no significant changes in respiratory rate. In vivo testing demonstrated comparable gas exchange parameters between SV-LCV and conventional ventilator across all cycles. Specifically, in cycle 1, the SV-LCV showed arterial blood gas (ABG) results of pH 7.54, PCO2 34.5 mmHg, and PO2 91.7 mmHg, compared to the control ventilator's ABG of pH 7.53, PCO2 37.1 mmHg, and PO2 134 mmHg. Cycle 2 exhibited ABG results of pH 7.53, PCO2 33.6 mmHg, and PO2 84.3 mmHg for SV-LCV, and pH 7.5, PCO2 34.2 mmHg, and PO2 93.5 mmHg for the control ventilator. Similarly, cycle 3 showed ABG results of pH 7.53, PCO2 32.1 mmHg, and PO2 127 mmHg for SV-LCV, and pH 7.5, PCO2 35.5 mmHg, and PO2 91.3 mmHg for the control ventilator.
conclusionThe SV-LCV provides similar gas exchange and respiratory mechanic profiles compared to a conventional ventilator. With a streamlined design and performance akin to commercially available ventilators, the SV-LCV presents a viable, readily available, and reliable short-term solution for overcoming ventilator supply shortages during crises.