Evidence map›Paper›PMID 38733400›Full record

ArticleEuropean spine journal : official publication of the European Spine Society, the European Spinal Deformity Society, and the European Section of the Cervical Spine Research Society2024

Endplate weakening during cage bed preparation significantly reduces endplate load capacity.

Anna-Katharina Calek, Frédéric Cornaz, Mauro Suter, Marie-Rosa Fasser, Mazda Farshad, Jonas Widmer

Abstract read
PubMed Publisher
In one paragraph

Article in European spine journal : official publication of the European Spine Society, the European Spinal Deformity Society, and the European Section of the Cervical Spine Research Society, 2024. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 10 papers, 1 of them a synthesis that pooled it.

0numbers the graph read from it
0cells of the map it votes in
10citing papers in PubMed, 1 pooled it
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

10 citing papers in PubMed, 1 synthesis or guideline pooled it.

  1. Pooled it
  2. Article
  3. Article
  4. Article
  5. Article
  6. Article
  7. Article
  8. Article
  9. Article
  10. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

6 authors.

Anna-Katharina CalekDepartment of Orthopedics, Balgrist University Hospital, University of Zurich, Forchstrasse 340, 8008, Zurich, Switzerland. anna-katharina.calek@balgrist.ch.ORCID 0000-0002-5499-5096
Frédéric CornazDepartment of Orthopedics, Balgrist University Hospital, University of Zurich, Forchstrasse 340, 8008, Zurich, Switzerland.
Mauro SuterSpine Biomechanics, Department of Orthopedic Surgery, Balgrist University Hospital, University of Zurich, Zurich, Switzerland.
Marie-Rosa FasserSpine Biomechanics, Department of Orthopedic Surgery, Balgrist University Hospital, University of Zurich, Zurich, Switzerland.
Mazda FarshadDepartment of Orthopedics, Balgrist University Hospital, University of Zurich, Forchstrasse 340, 8008, Zurich, Switzerland.
Jonas WidmerSpine Biomechanics, Department of Orthopedic Surgery, Balgrist University Hospital, University of Zurich, Zurich, Switzerland.

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

purposeTo analyze the effect of endplate weakness prior to PLIF or TLIF cage implantation and compare it to the opposite intact endplate of the same vertebral body. In addition, the influence of bone quality on endplate resistance was investigated.

methodsTwenty-two human lumbar vertebrae were tested in a ramp-to-failure test. One endplate of each vertebral body was tested intact and the other after weakening with a rasp (over an area of 200 mm

resultsWith an intact endplate and a TLIF cage, the median force to failure was 1276.3N (693.1-1980.6N). Endplate weakening reduced axial endplate resistance to failure by 15% (0-23%). With an intact endplate and a PLIF cage, the median force to failure was 1057.2N (701.2-1735.5N). Endplate weakening reduced axial endplate resistance to failure by 36.6% (7-47.9%). Bone quality correlated linearly with the force at which endplate failure occurred. Intact and weakened endplates showed a strong positive correlation: intact-TLIF: r = 0.964, slope of the regression line (slope) = 11.8, p < 0.001; intact-PLIF: r = 0.909, slope = 11.2, p = 5.5E-05; weakened-TLIF: r = 0.973, slope = 12.5, p < 0.001; weakened-PLIF: r = 0.836, slope = 6, p = 0.003.

conclusionWeakening of the endplate during cage bed preparation significantly reduces the resistance of the endplate to subsidence to failure: endplate load capacity is reduced by 15% with TLIF and 37% with PLIF. Bone quality correlates with the force at which endplate failure occurs.

Indexed as

Lumbar VertebraeSpinal FusionWeight-BearingAdultAgedAged, 80 and overBiomechanical PhenomenaFemaleHumansMaleMiddle AgedBiomechanical studyCageDorsal fusionEndplatePLIFSubsidenceTLIF

Identifiers

What OpenQuestion holds

Textmetadata
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.