ArticleUltrasound in medicine & biology2024
Color Flow Ultrasound Spatial Sampling Beam Density for Partial Volume-Corrected Three-Dimensional Volume Flow (3DVF): Theory, Simulation, and Experiment.
Article in Ultrasound in medicine & biology, 2024. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 2 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
2 citing papers in PubMed.
- Forward-Viewing Intravascular Ultrasound: Design and Fabrication for High Sensitivity 3D Imaging of Hemodynamics.IEEE transactions on bio-medical engineering · 2026Article
- Intravascular ultrasound wall shear stress imaging in stented coronary arteries with ultrafast Doppler.Scientific reports · 2026Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
5 authors.
Funding
Abstract
objectiveThe purpose of this study was to quantify the accuracy of partial volume-corrected three-dimensional volume flow (3DVF) measurements as a function of spatial sampling beam density using carefully-designed parametric analyses in order to inform the target applications of 3DVF.
methodsExperimental investigations employed a mechanically-swept curvilinear ultrasound array to acquire 3D color flow (6.3 MHz) images in flow phantoms consisting of four lumen diameters (6.35, 4.88, 3.18 and 1.65 mm) with volume flow rates of 440, 260, 110 and 30 mL/min, respectively. Partial volume-corrected three-dimensional volume flow (3DVF) measurements, based on the Gaussian surface integration principle, were computed at five regions of interest positioned between depths of 2 and 6 cm in 1 cm increments. At each depth, the color flow beam point spread function (PSF) was also determined, using in-phase/quadrature data, such that 3DVF bias could then be related to spatial sampling beam density. Corresponding simulations were performed for a laminar parabolic flow profile that was sampled using the experimentally-measured PSFs. Volume flow was computed for all combinations of lumen diameters and the PSFs at each depth.
resultsAccurate 3DVF measurements, i.e., bias less than ±20%, were achieved for spatial sampling beam densities where at least 6 elevational color flow beams could be positioned across the lumen. In these cases, greater than 8 lateral color flow beams were present. PSF measurements showed an average lateral-to-elevational beam width asymmetry of 1:2. Volume flow measurement bias increased as the color flow beam spatial sampling density within the lumen decreased.
conclusionApplications of 3DVF, particularly those in the clinical domain, should focus on areas where a spatial sampling density of 6 × 6 (lateral x elevational) beams can be realized in order to minimize measurement bias. Matrix-based ultrasound arrays that possess symmetric PSFs may be advantageous to achieve adequate beam densities in smaller vessels.