ArticleCritical care explorations2024
The Presence of Blood in a Strain Gauge Pressure Transducer Has a Clinical Effect on the Accuracy of Intracranial Pressure Readings.
Article in Critical care explorations, 2024. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 2 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
2 citing papers in PubMed.
- Intracranial Pressure Estimation Using B4C Device: A Systematic Review on the Validation Studies.Neurocritical care · 2026Review
- External Ventricular Drain Column Height Fails to Explain Cerebrospinal Fluid Drainage Volumes.Nursing in critical care · 2025Observational
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
5 authors.
Funding
Abstract
importancePatients admitted with cerebral hemorrhage or cerebral edema often undergo external ventricular drain (EVD) placement to monitor and manage intracranial pressure (ICP). A strain gauge transducer accompanies the EVD to convert a pressure signal to an electrical waveform and assign a numeric value to the ICP.
objectivesThis study explored ICP accuracy in the presence of blood and other viscous fluid contaminates in the transducer.
designPreclinical comparative design study.
settingLaboratory setting using two Natus EVDs, two strain gauge transducers, and a sealed pressure chamber.
participantsNo human subjects or animal models were used.
interventionsA control transducer primed with saline was compared with an investigational transducer primed with blood or with saline/glycerol mixtures in mass:mass ratios of 25%, 50%, 75%, and 100% glycerol. Volume in a sealed chamber was manipulated to reflect changes in ICP to explore the impact of contaminates on pressure measurement. MEASUREMENTS AND MAIN
resultsFrom 90 paired observations, ICP readings were statistically significantly different between the control (saline) and experimental (glycerol or blood) transducers. The time to a stable pressure reading was significantly different for saline vs. 25% glycerol (< 0.0005), 50% glycerol (< 0.005), 75% glycerol (< 0.0001), 100% glycerol (< 0.0005), and blood (< 0.0005). A difference in resting stable pressure was observed for saline vs. blood primed transducers (0.041). CONCLUSIONS AND RELEVANCE: There are statistically significant and clinically relevant differences in time to a stable pressure reading when contaminates are introduced into a closed drainage system. Changing a transducer based on the presence of blood contaminate should be considered to improve accuracy but must be weighed against the risk of introducing infection.