Evidence map›Paper›PMID 38699670›Full record

ArticleAdvances in radiation oncology2024

Commissioning and Assessment of Radiation Field and Dose Inhomogeneity for a Dual X-ray Tube Cabinet Irradiator: To Ensure Accurate Dosimetry in Radiation Biology Experiments.

Mitchell Polizzi, Kristoffer Valerie, Siyong Kim

Abstract read
In one paragraph

Article in Advances in radiation oncology, 2024. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 1 paper.

0numbers the graph read from it
0cells of the map it votes in
1citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

1 citing paper in PubMed.

  1. Review
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

3 authors.

Mitchell PolizziDepartment of Radiation Oncology, Virginia Commonwealth University, Richmond, Virginia.
Kristoffer ValerieDepartment of Radiation Oncology, Virginia Commonwealth University, Richmond, Virginia.
Siyong KimDepartment of Radiation Oncology, Virginia Commonwealth University, Richmond, Virginia.

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

Purpose: Standardization of x-ray cabinet irradiator dose, geometry, and calibration reporting is an ongoing process. Multi-tube designs have been introduced into the preclinical market and give a theoretical benefit but have not been widely assessed for use in preclinical irradiation conditions. The aim of this study was to report our experience commissioning a dual x-ray source cabinet irradiator (CIXD, Xstrahl Limited, United Kingdom) and assess the dose distribution for various experimental conditions. Methods and Materials: Half-value layer (HVL) measurement, profile measurements, and output calibration were performed using a calibrated ion chamber. Constancy measurements were performed twice daily over 2 weeks to assess output fluctuations. Film measurements were completed using solid water to assess percent depth dose and homogeneity within the field and within variable thicknesses of solid water and phosphate-buffered saline solution. Film measurements were repeated for various arrangements of petri dishes filled with phosphate-buffered saline or water and in a 3D-printed mouse phantom. Results: The x-ray tubes had a measured in-air output of 1.27 Gy/min. The HVL was 1.7 mm Cu. The upper and lower tubes both exhibited the heel effect, but when operated simultaneously, the effect was reduced. Ion chamber measurements revealed a 15% dose inhomogeneity within the tray area (18 × 18 cm Conclusions: X-ray cell culture and animal irradiation with dual tube cabinet irradiation is efficient and robust when using established dosimetric tools to confirm output and homogeneity. The conditions assumed for calibrations are often not maintained during experiments. We have confirmed that inhomogeneities are present for single-tube use; however, they are reduced with simultaneous tube use. Additional dosimetric monitoring should be performed for each unique irradiation setup.

Identifiers

PMID38699670
PMCPMC11063221

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.