Trial reportBMC pulmonary medicine2024
The effects of real-time waveform analysis software on patient ventilator synchronization during pressure support ventilation: a randomized crossover physiological study.
Trial report in BMC pulmonary medicine, 2024. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 4 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
4 citing papers in PubMed, 10 citations in OpenAlex.
- Closed-loop synchronization versus conventional synchronization in spontaneously breathing pediatric patients (CHESTSIPP) - a randomized controlled cross-over study.Frontiers in medicine · 2026Trial
- Clinical and ventilator waveforms of Chinese patients during pressure support ventilation.Scientific data · 2025Article
- Ability to identify patient-ventilator asynchronies in intensive care unit professionals: A multicenter cross-sectional analytical study.Journal of critical care medicine (Universitatea de Medicina si Farmacie din Targu-Mures) · 2025Article
- Development and performance assessment of a novel scroll compressor-based oxygen generator integrated ventilator.Scientific reports · 2025Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
3 authors at 2 institutions in 1 country.
Funding
No grant is acknowledged in the PubMed record.
Abstract
backgroundPatient-ventilator asynchrony commonly occurs during pressure support ventilation (PSV). IntelliSync + software (Hamilton Medical AG, Bonaduz, Switzerland) is a new ventilation technology that continuously analyzes ventilator waveforms to detect the beginning and end of patient inspiration in real time. This study aimed to evaluate the physiological effect of IntelliSync + software on inspiratory trigger delay time, delta airway (P
methodsA randomized crossover physiologic study was conducted in 14 mechanically ventilated patients under PSV. Patients were randomly assigned to receive conventional flow trigger and cycling, inspiratory trigger synchronization (I-sync), cycle synchronization (C-sync), and inspiratory trigger and cycle synchronization (I/C-sync) for 15 min at each step. Other ventilator settings were kept constant. P
resultsThe time to initiate the trigger was significantly shorter with I-sync compared to baseline (208.9±91.7 vs. 301.4±131.7 msec; P = 0.002) and I/C-sync compared to baseline (222.8±94.0 vs. 301.4±131.7 msec; P = 0.005). The I/C-sync group had significantly lower delta P
conclusionsIntelliSync + software reduced inspiratory trigger delay time compared to the conventional flow trigger system during PSV mode. However, no significant improvements in cycle delay time and other physiological variables were observed with IntelliSync + software.
trial registrationThis study was registered in the Thai Clinical Trial Registry (TCTR20200528003; date of registration 28/05/2020).