Evidence map›Paper›PMID 38693506›Full record

Trial reportBMC pulmonary medicine2024

The effects of real-time waveform analysis software on patient ventilator synchronization during pressure support ventilation: a randomized crossover physiological study.

Barnpot Nakornnoi, Jamsak Tscheikuna, Nuttapol Rittayamai

Open access · goldAbstract readRandomized Controlled Trial
In one paragraph

Trial report in BMC pulmonary medicine, 2024. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 4 papers.

0numbers the graph read from it
0cells of the map it votes in
4citing papers in PubMed
4.2field-weighted citation impact, top 6% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

4 citing papers in PubMed, 10 citations in OpenAlex.

  1. Trial
  2. Article
  3. Ability to identify patient-ventilator asynchronies in intensive care unit professionals: A multicenter cross-sectional analytical study.Journal of critical care medicine (Universitatea de Medicina si Farmacie din Targu-Mures) · 2025
    Article
  4. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

3 authors at 2 institutions in 1 country.

Barnpot NakornnoiDivision of Respiratory Diseases and Tuberculosis, Department of Medicine, Faculty of Medicine Siriraj Hospital, Mahidol University, Bangkok, Thailand.
Jamsak TscheikunaDivision of Respiratory Diseases and Tuberculosis, Department of Medicine, Faculty of Medicine Siriraj Hospital, Mahidol University, Bangkok, Thailand.
Nuttapol RittayamaiDivision of Respiratory Diseases and Tuberculosis, Department of Medicine, Faculty of Medicine Siriraj Hospital, Mahidol University, Bangkok, Thailand. nuttapol.rit@mahidol.ac.th.
Mahidol University · THSiriraj Hospital · TH

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

backgroundPatient-ventilator asynchrony commonly occurs during pressure support ventilation (PSV). IntelliSync + software (Hamilton Medical AG, Bonaduz, Switzerland) is a new ventilation technology that continuously analyzes ventilator waveforms to detect the beginning and end of patient inspiration in real time. This study aimed to evaluate the physiological effect of IntelliSync + software on inspiratory trigger delay time, delta airway (P

methodsA randomized crossover physiologic study was conducted in 14 mechanically ventilated patients under PSV. Patients were randomly assigned to receive conventional flow trigger and cycling, inspiratory trigger synchronization (I-sync), cycle synchronization (C-sync), and inspiratory trigger and cycle synchronization (I/C-sync) for 15 min at each step. Other ventilator settings were kept constant. P

resultsThe time to initiate the trigger was significantly shorter with I-sync compared to baseline (208.9±91.7 vs. 301.4±131.7 msec; P = 0.002) and I/C-sync compared to baseline (222.8±94.0 vs. 301.4±131.7 msec; P = 0.005). The I/C-sync group had significantly lower delta P

conclusionsIntelliSync + software reduced inspiratory trigger delay time compared to the conventional flow trigger system during PSV mode. However, no significant improvements in cycle delay time and other physiological variables were observed with IntelliSync + software.

trial registrationThis study was registered in the Thai Clinical Trial Registry (TCTR20200528003; date of registration 28/05/2020).

Indexed as

Cross-Over StudiesSoftwareAgedFemaleHemodynamicsHumansMaleMiddle AgedPositive-Pressure RespirationRespiration, ArtificialRespiratory RateAsynchronyEsophageal pressurePatient-ventilator interactionPressure support ventilation

Identifiers

PMID38693506
PMCPMC11064376
OpenAlexW4396561843

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.