ArticleJournal of applied clinical medical physics2024
The impact of automatic tube current modulation related settings of a modern GE CT scanner on image quality and patient dose; details do matter.
Article in Journal of applied clinical medical physics, 2024. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 8 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
8 citing papers in PubMed, 7 citations in OpenAlex.
- Assessment of image quality and detectability in computed tomography: Inter-laboratory comparison of Fourier-based metrics.Journal of applied clinical medical physics · 2026Article
- Relationship between sacroiliac joint involvement and intestinal disease activity in newly diagnosed Crohn's disease.BMC gastroenterology · 2026Article
- Comparison of deep learning reconstruction algorithms to improve image quality of dual-energy carotid CT angiography under dual-low scan.Neuroradiology · 2026Article
- Attenuation-based ultra-low-dose lung computed tomography at 0.1 mSv to 0.3 mSv effective dose in children.Pediatric radiology · 2026Article
- Improved organ absorbed dose estimation in abdominal CT using DICOM header-based tube current modulation profiles: Validation with measurements and Monte Carlo simulations.Journal of applied clinical medical physics · 2025Article
- The impact of automatic tube current modulation related settings of a modern GE CT scanner on image quality and patient dose; details do matter.Journal of applied clinical medical physics · 2024Article
- Artificial Intelligence in Chest Computed Tomography Optimization: A Task-based Framework for Balancing Detectability and Radiation Dose.Journal of medical physicsReview
- Optimizing Chest Computed Tomography Imaging Protocols: A Narrative Review on Dose Reduction and Diagnostic Efficacy.Journal of medical physicsReview
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
3 authors at 1 institution in 1 country.
Funding
No grant is acknowledged in the PubMed record.
Abstract
purposeTo investigate the operation principles of the automatic tube current modulation (ATCM) of a modern GE healthcare CT scanner, and the impact of related settings on image quality and patient dose. MATERIAL &
methodsA dedicated phantom (Mercury 4.0) was scanned using two of the most frequently used clinical scanning protocols (chest and abdomen-pelvis). The preset protocol settings were used as starting points (reference conditions). Scan direction, scan mode (helical vs. axial), total beam width, tube potential (kVp), and ATCM settings were then modified individually to understand their impact on radiation dose and image quality. Regarding the ATCM settings, the SmartmA minimum and maximum mA limits, and the noise index (NI) values were varied. As surrogates of patient dose, the CTDI
resultsThe scanning direction did not have any effect on ATCM curves, unlike what has been observed in CT scanners from other manufacturers. Total beam width does matter, however, the SmartmA limit settings and kVp selection had the greatest impact on image quality and dose. It was seen that improper minimum mA limit settings practically invalidated the ATCM operation. In contrast, when full modulation was allowed without restrictions, noise standard deviation, and detectability index became much more consistent across the wide range of phantom diameters. For lower kVp settings an impressive dose reduction was observed that requires further investigation.
conclusionSmartmA is a tool that if not properly used may increase the patient doses considerably. Therefore, its settings should be carefully adjusted for each preset different clinical protocol.