Evidence map›Paper›PMID 38655007›Full record

ArticleJOR spine2024

Comparison of four in vitro test methods to assess nucleus pulposus replacement device expulsion risk.

Tamanna Rahman, Matthew J Kibble, Gianluca Harbert, Nigel Smith, Erik Brewer, Thomas P Schaer, Nicolas Newell

Open access · goldAbstract read
In one paragraph

Article in JOR spine, 2024. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 4 papers, 1 of them a synthesis that pooled it.

0numbers the graph read from it
0cells of the map it votes in
4citing papers in PubMed, 1 pooled it
1.3field-weighted citation impact, top 22% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

4 citing papers in PubMed, 1 synthesis or guideline pooled it, 3 citations in OpenAlex.

  1. Lumbar disc herniation modelling: a review of ex-vivo mechanical models and a comparison with clinical data.European spine journal : official publication of the European Spine Society, the European Spinal Deformity Society, and the European Section of the Cervical Spine Research Society · 2025
    Pooled it
  2. Article
  3. Article
  4. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

7 authors at 4 institutions in 2 countries.

Tamanna RahmanDepartment of Bioengineering Imperial College London London UK.ORCID https://orcid.org/0000-0002-0331-0288
Matthew J KibbleDepartment of Bioengineering Imperial College London London UK.
Gianluca HarbertDepartment of Bioengineering Imperial College London London UK.
Nigel SmithDivision of Surgery and Interventional Science University College London Stanmore UK.
Erik BrewerDepartment of Biomedical Engineering Rowan University Glassboro New Jersey USA.
Thomas P SchaerDepartment of Clinical Studies New Bolton Center University of Pennsylvania School of Veterinary Medicine Kennett Square Pennsylvania USA.
Nicolas NewellDepartment of Bioengineering Imperial College London London UK.ORCID https://orcid.org/0000-0003-1288-3990
Imperial College London · GBRowan University · USStanmore College · GBUniversity of Pennsylvania · US

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

Background: Nucleus replacement devices (NRDs) are not routinely used in clinic, predominantly due to the risk of device expulsion. Rigorous in vitro testing may enable failure mechanisms to be identified prior to clinical trials; however, current testing standards do not specify a particular expulsion test. Multiple methods have therefore been developed, complicating comparisons between NRD designs. Thus, this study assessed the effectiveness of four previously reported expulsion testing protocols; hula-hoop (Protocol 1), adapted hula-hoop (Protocol 2), eccentric cycling (Protocol 3), and ramp to failure (Protocol 4), applied to two NRDs, one preformed and one in situ curing. Methods: Nucleus material was removed from 40 bovine tail intervertebral disks. A NRD was inserted posteriorly into each cavity and the disks were subjected to one of four expulsion protocols. Results: NRD response was dependent on both the NRD design and the loading protocol. Protocol 1 resulted in higher migration and earlier failure rates compared to Protocol 2 in both NRDs. The preformed NRD was more likely to migrate when protocols incorporated rotation. The NRDs had equal migration (60%) and expulsion (60%) rates when using unilateral bending and ramp testing. Combining the results of multiple tests revealed complimentary information regarding the NRD response. Conclusions: Adapted hula-hoop (Protocol 2) and ramp to failure (Protocol 4), combined with fluoroscopic analysis, revealed complimentary insights regarding migration and failure risk. Therefore, when adopting the surgical approach and animal model used in this study, it is recommended that NRD performance be assessed using both a cyclic and ramp loading protocol.

Indexed as

biomechanical testingexpulsionhydrogelintervertebral disknucleus replacementspine

Identifiers

PMID38655007
PMCPMC11037461
OpenAlexW4395031596

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.