Real-Time Temperature-Controlled Retinal Laser Irradiation in Rabbits.

Claus von der Burchard, Christopher Kren, Jan-Erik Fleger, Dirk Theisen-Kunde, Veit Danicke, Hossam S Abbas, Viktoria Kleyman, Johann Roider, Ralf Brinkmann

Open access · goldAbstract read
In one paragraph

Article in Translational vision science & technology, 2024. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 3 papers.

0numbers the graph read from it
0cells of the map it votes in
3citing papers in PubMed
0.7field-weighted citation impact, top 32% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

3 citing papers in PubMed, 3 citations in OpenAlex.

  1. Article
  2. Article
  3. Model Predictive Temperature Control for Retinal Laser Treatments.Translational vision science & technology · 2024
    Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

9 authors at 4 institutions in 1 country.

Claus von der BurchardDepartment of Ophthalmology, Christian-Albrechts-University of Kiel, Kiel, Germany.
Christopher KrenMedical Laser Center Lübeck, Lübeck, Germany.
Jan-Erik FlegerDepartment of Ophthalmology, Christian-Albrechts-University of Kiel, Kiel, Germany.
Dirk Theisen-KundeMedical Laser Center Lübeck, Lübeck, Germany.
Veit DanickeMedical Laser Center Lübeck, Lübeck, Germany.
Hossam S AbbasMedical Laser Center Lübeck, Lübeck, Germany.
Viktoria KleymanInstitute of Automatic Control, Leibniz University Hannover, Hannover, Germany.
Johann RoiderDepartment of Ophthalmology, Christian-Albrechts-University of Kiel, Kiel, Germany.
Ralf BrinkmannMedical Laser Center Lübeck, Lübeck, Germany.
Christian-Albrechts-Universität zu Kiel · DEMedizinisches Laserzentrum Lübeck (Germany) · DEUniversity of Lübeck · DELeibniz University Hannover · DE

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

Purpose: Subdamaging thermal retinal laser therapy has the potential to induce regenerative stimuli in retinal diseases, but validated dosimetry is missing. Real-time optoacoustic temperature determination and control could close this gap. This study investigates a first in vivo application. Methods: Two iterations of a control module that were optically coupled in between a continuous-wave commercial laser source and a commercial slit lamp were evaluated on chinchilla rabbits. The module allows extraction of the temperature rise in real time and can control the power of the therapy laser such that a predefined temperature rise at the retina is quickly achieved and held constant. Irradiations with aim temperatures from 45°C to 69°C were performed on a diameter of 200 µm and a heating time of 100 ms. Results: We analyzed 424 temperature-guided irradiations in nine eyes of five rabbits. The mean difference between the measured and aim temperature was -0.04°C ± 0.98°C. The following ED50 values for visibility thresholds could be determined: 58.6°C for funduscopic visibility, 57.7°C for fluorescein angiography, and 57.0°C for OCT. In all measurements, the correlation of tissue effect was higher to the temperature than to the average heating laser power used. Conclusions: The system was able to reliably perform temperature-guided irradiations, which allowed for better tissue effect control than simple power control. This approach could enhance the accuracy, safety, and reproducibility of thermal stimulating laser therapy. Translational Relevance: This study is a bridge between preclinical ex vivo experiments and a pilot clinical study.

Indexed as

RetinaRetinal DiseasesAnimalsFluorescein AngiographyRabbitsReproducibility of ResultsTemperature

Identifiers

PMID38639930
PMCPMC11037498
OpenAlexW4394952452

What OpenQuestion holds

Textmetadata
LicenceCC BY-NC-ND
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.