Evidence map›Paper›PMID 38634799›Full record

ArticleJournal of applied clinical medical physics2024

Evaluation and mitigation of deformable image registration uncertainties for MRI-guided adaptive radiotherapy.

Hualiang Zhong, Kristofer K Kainz, Eric S Paulson

Abstract readEvaluation Study
In one paragraph

Article in Journal of applied clinical medical physics, 2024. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 4 papers.

0numbers the graph read from it
0cells of the map it votes in
4citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

4 citing papers in PubMed.

  1. Article
  2. Article
  3. Deformable dose accumulation is required for adaptive radiotherapy practice.Journal of applied clinical medical physics · 2024
    Article
  4. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

3 authors.

Hualiang ZhongDepartment of Radiation Oncology, Medical College of Wisconsin, Milwaukee, Wisconsin, USA.ORCID https://orcid.org/0000-0002-3519-0815
Kristofer K KainzDepartment of Radiation Oncology, Medical College of Wisconsin, Milwaukee, Wisconsin, USA.
Eric S PaulsonDepartment of Radiation Oncology, Medical College of Wisconsin, Milwaukee, Wisconsin, USA.

Funding

eXtended Modular ANthropomorphic (XMAN) phantom for Imaging and Treatment Optimization in Radiotherapy.R01EB032680 · NIBIB · UNIVERSITY OF MARYLAND BALTIMORE · PI REN, LEI, YANG, XIAOFENG · 2022 to 2025
$2.4M
NIBIB NIH HHS R01 EB032680NIH
6 · The paper itself

Abstract

purposeWe evaluate the performance of a deformable image registration (DIR) software package in registering abdominal magnetic resonance images (MRIs) and then develop a mechanical modeling method to mitigate detected DIR uncertainties. MATERIALS AND

methodsThree evaluation metrics, namely mean displacement to agreement (MDA), DICE similarity coefficient (DSC), and standard deviation of Jacobian determinants (STD-JD), are used to assess the multi-modality (MM), contour-consistency (CC), and image-intensity (II)-based DIR algorithms in the MIM software package, as well as an in-house developed, contour matching-based finite element method (CM-FEM). Furthermore, we develop a hybrid FEM registration technique to modify the displacement vector field of each MIM registration. The MIM and FEM registrations were evaluated on MRIs obtained from 10 abdominal cancer patients. One-tailed Wilcoxon-Mann-Whitney (WMW) tests were conducted to compare the MIM registrations with their FEM modifications.

resultsFor the registrations performed with the MIM-CC, MIM-MM, MIM-II, and CM-FEM algorithms, their average MDAs are 0.62 ± 0.27, 2.39 ± 1.30, 3.07 ± 2.42, 1.04 ± 0.72 mm, and average DSCs are 0.94 ± 0.03, 0.80 ± 0.12, 0.77 ± 0.15, 0.90 ± 0.11, respectively. The p-values of the WMW tests between the MIM registrations and their FEM modifications are less than 0.0084 for STD-JDs and greater than 0.87 for MDA and DSC.

conclusionsAmong the three MIM DIR algorithms, MIM-CC shows the smallest errors in terms of MDA and DSC but exhibits significant Jacobian uncertainties in the interior regions of abdominal organs. The hybrid FEM technique effectively mitigates the Jacobian uncertainties in these regions.

Indexed as

AlgorithmsImage Processing, Computer-AssistedMagnetic Resonance ImagingRadiotherapy, Image-GuidedRadiotherapy Planning, Computer-AssistedAbdominal NeoplasmsHumansRadiotherapy DosageRadiotherapy, Intensity-ModulatedSoftwareUncertaintyadaptive radiotherapydeformable image registrationevaluation metricsmagnetic resonance imageuncertainty reduction

Identifiers

PMID38634799
PMCPMC11163488

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.