Evidence map›Paper›PMID 38626156›Full record

ArticlePloS one2024

Reliability and performance of the IRRAflow® system for intracranial lavage and evacuation of hematomas-A technical note.

Mette Haldrup, Mojtaba Nazari, Chenghao Gu, Mads Rasmussen, Stig Dyrskog, Claus Ziegler Simonsen, Mads Grønhøj, Frantz Rom Poulsen, Naveed Ur Rehman, Anders Rosendal Korshoej

Abstract read
In one paragraph

Article in PloS one, 2024. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 4 papers, 1 of them a synthesis that pooled it.

0numbers the graph read from it
0cells of the map it votes in
4citing papers in PubMed, 1 pooled it
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

4 citing papers in PubMed, 1 synthesis or guideline pooled it.

  1. Pooled it
  2. Article
  3. Article
  4. Continuous Ventricular Irrigation for Intraventricular Hemorrhage.Current neurology and neuroscience reports · 2025
    Review
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

10 authors.

Mette HaldrupDepartment of Neurosurgery, Aarhus University Hospital, Aarhus N, Denmark.ORCID 0009-0008-7173-3160
Mojtaba NazariDepartment of Engineering, Electrical and Computer Engineering, Aarhus University, Aarhus N, Denmark.
Chenghao GuDepartment of Neurosurgery, Aarhus University Hospital, Aarhus N, Denmark.
Mads RasmussenDepartment of Clinical Medicine, Aarhus University, Aarhus N, Denmark.
Stig DyrskogDepartment of Intensive Care, Aarhus University Hospital, Aarhus N, Denmark.
Claus Ziegler SimonsenDepartment of Intensive Care, Aarhus University Hospital, Aarhus N, Denmark.
Mads GrønhøjDepartment of Neurosurgery, Odense University Hospital, Odense, Denmark.
Frantz Rom PoulsenDepartment of Neurosurgery, Odense University Hospital, Odense, Denmark.ORCID 0000-0001-5715-6901
Naveed Ur RehmanDepartment of Engineering, Electrical and Computer Engineering, Aarhus University, Aarhus N, Denmark.
Anders Rosendal KorshoejDepartment of Neurosurgery, Aarhus University Hospital, Aarhus N, Denmark.ORCID 0000-0003-4285-8171

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

backgroundIntraventricular hemorrhage (IVH) is a severe condition with poor outcomes and high mortality. IRRAflow® (IRRAS AB) is a new technology introduced to accelerate IVH clearance by minimally invasive wash-out. The IRRAflow® system performs active and controlled intracranial irrigation and aspiration with physiological saline, while simultaneously monitoring and maintaining a stable intracranial pressure (ICP). We addressed important aspects of the device implementation and intracranial lavage.

methodTo allow versatile investigation of multiple device parameters, we designed an ex vivo lab setup. We evaluated 1) compatibility between the IRRAflow® catheter and the Silverline f10 bolt (Spiegelberg), 2) the physiological and hydrodynamic effects of varying the IRRAflow® settings, 3) the accuracy of the IRRAflow® injection volumes, and 4) the reliability of the internal ICP monitor of the IRRAflow®.

resultsThe IRRAflow® catheter was not compatible with Silverline bolt fixation, which was associated with leakage and obstruction. Design space exploration of IRRAflow® settings revealed that appropriate settings included irrigation rate 20 ml/h with a drainage bag height at 0 cm, irrigation rate 90 ml/h with a drainage bag height at 19 cm and irrigation rate 180 ml/h with a drainage bag height at 29 cm. We found the injection volume performed by the IRRAflow® to be stable and reliable, while the internal ICP monitor was compromised in several ways. We observed a significant mean drift difference of 3.16 mmHg (variance 0.4, p = 0.05) over a 24-hour test period with a mean 24-hour drift of 3.66 mmHg (variance 0.28) in the pressures measured by the IRRAflow® compared to 0.5 mmHg (variance 1.12) in the Raumedic measured pressures.

conclusionBolting of the IRRAflow® catheter using the Medtronic Silverline® bolt is not recommendable. Increased irrigation rates are recommendable followed by a decrease in drainage bag level. ICP measurement using the IRRAflow® device was unreliable and should be accompanied by a control ICP monitor device in clinical settings.

Indexed as

Intracranial PressureTherapeutic IrrigationCerebral HemorrhageHematomaHumansMonitoring, PhysiologicReproducibility of Results

Identifiers

PMID38626156
PMCPMC11020765

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

None linked