ArticleInternational journal of computer assisted radiology and surgery2024
The development and testing of a smart sensorized guide wire for catheterization in a "blood" vessel phantom to support aortic valve implementation.
Article in International journal of computer assisted radiology and surgery, 2024. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Not yet cited in PubMed.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
0 citing papers in PubMed, 0 citations in OpenAlex.
No citing paper in PubMed yet.
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
5 authors at 2 institutions in 1 country.
Funding
No grant is acknowledged in the PubMed record.
Abstract
purposeHeart valve disease is commonly treated by minimally invasive procedures with guide wires and catheterization. The main purpose of this study is to find out whether an extension of the guide wire with a sensor can support the surgeon within the blood vessel to reduce X-ray necessity.
methodsA smart guide wire is developed by an extension with a flex-bending sensor to evaluate the sensor signal with and without "blood" flow at a constant compression force. Various surgically relevant investigations are performed. For assessment, the mean temporal average of the moving averaged filtered ADC signal and a subsequent FFT are carried out.
resultsResults show that there is a smaller sensor signal when the applied force or bending at the sensor is higher. In all investigations, there was a different sensor signal. The flex-bending sensor can detect the effect of pulsatile flow. The smallest temporal averaged signal difference between reference and clamp in the front wire's tip is 1.09%. For example, the mean temporal average of the filtered ADC signal for different clinically relevant scenarios is between 2550 and 2900.
conclusionsThe results show that the sensorized guide wire developed for catheterization can support aortic valve implementation. The sensor sensitivity is sufficient to detect even very small variations within the blood vessel and therefore is promising to support catheterization heart valve surgeries in future.