The dilution evaluation as a corrective measure for interference in the white blood cell scattergram in Beckman Coulter DxH 900.

Filipa P Freitas, Jorge Reis, Joana Oliveira, Pedro Mota Veiga, Ana Raquel Paiva, Rui Soares

Abstract read
In one paragraph

Article in Journal of clinical laboratory analysis, 2024. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 2 papers.

0numbers the graph read from it
0cells of the map it votes in
2citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

2 citing papers in PubMed.

  1. Article
  2. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

6 authors.

Filipa P FreitasClinical Pathology Service, Portuguese Institute of Oncology of Coimbra Francisco Gentil, EPE, Coimbra, Portugal.
Jorge ReisClinical Pathology Service, Portuguese Institute of Oncology of Coimbra Francisco Gentil, EPE, Coimbra, Portugal.
Joana OliveiraClinical Pathology Service, Portuguese Institute of Oncology of Coimbra Francisco Gentil, EPE, Coimbra, Portugal.ORCID https://orcid.org/0000-0002-3826-3610
Pedro Mota VeigaUniversity of Maia, Maia, Portugal.
Ana Raquel PaivaClinical Pathology Service, Portuguese Institute of Oncology of Coimbra Francisco Gentil, EPE, Coimbra, Portugal.
Rui SoaresClinical Pathology Service, Portuguese Institute of Oncology of Coimbra Francisco Gentil, EPE, Coimbra, Portugal.

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

backgroundThe Beckman Coulter DxH 900 is a haematological analyser capable of counting and sizing blood cells, and obtaining a complete blood cell count (CBC). This analyses different parameters of red blood cells (RBC), platelets and white blood cells/leukocytes. Some automated CBC counters present limitations due to specimen characteristics, abnormal cells or both factors. In the presence of abnormalities, the DxH 900 has a flagging system, warning the laboratory technician that something needs to be verified. In the present work, we evaluated samples from oncologic patients, presenting a population erroneously perceived as being lymphocytes. The most common explanations for this situation are RBC resistant to lysis or serum hyperbilirubinaemia.

methodsIn an attempt to solve and understand what the cause of this problem might be, we diluted our samples (1:3) and analysed the serum total bilirubin. To identify cells' abnormalities, the samples were also analysed by manual DLC counts. During the study, we also checked the different flags presented by the equipment.

resultsThe results evidenced that the major interference was due to RBC lysis resistance, corresponding to 94.7% of the cases, while hyperbilirubinaemia was only present in 73.4%. Besides, we determined that some samples with normal bilirubin levels also presented interference, suggesting that hyperbilirubinaemia was not the main cause of the error. The most recurrent flag observed was "High event rate".

conclusionThe dilution solved all of the observed interferences. The results between diluted and manual counts showed a strong correlation, leading us to introduce dilution in our laboratory routine.

Indexed as

LeukocytesBilirubinHumansLeukocyte CountBilirubindilutionhigh event ratehyperbilirubinaemiared blood cell lysis resistancewhite blood cell scattergram

Identifiers

PMID38594837
PMCPMC11073809

What OpenQuestion holds

Textmetadata
LicenceCC BY-NC
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.