The dilution evaluation as a corrective measure for interference in the white blood cell scattergram in Beckman Coulter DxH 900.
Article in Journal of clinical laboratory analysis, 2024. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 2 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
2 citing papers in PubMed.
- Evaluation of pre-dilution combined with optical/fluorescent platelet counting for correcting pseudothrombocytopenia.Frontiers in medicine · 2026Article
- The dilution evaluation as a corrective measure for interference in the white blood cell scattergram in Beckman Coulter DxH 900.Journal of clinical laboratory analysis · 2024Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
6 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
backgroundThe Beckman Coulter DxH 900 is a haematological analyser capable of counting and sizing blood cells, and obtaining a complete blood cell count (CBC). This analyses different parameters of red blood cells (RBC), platelets and white blood cells/leukocytes. Some automated CBC counters present limitations due to specimen characteristics, abnormal cells or both factors. In the presence of abnormalities, the DxH 900 has a flagging system, warning the laboratory technician that something needs to be verified. In the present work, we evaluated samples from oncologic patients, presenting a population erroneously perceived as being lymphocytes. The most common explanations for this situation are RBC resistant to lysis or serum hyperbilirubinaemia.
methodsIn an attempt to solve and understand what the cause of this problem might be, we diluted our samples (1:3) and analysed the serum total bilirubin. To identify cells' abnormalities, the samples were also analysed by manual DLC counts. During the study, we also checked the different flags presented by the equipment.
resultsThe results evidenced that the major interference was due to RBC lysis resistance, corresponding to 94.7% of the cases, while hyperbilirubinaemia was only present in 73.4%. Besides, we determined that some samples with normal bilirubin levels also presented interference, suggesting that hyperbilirubinaemia was not the main cause of the error. The most recurrent flag observed was "High event rate".
conclusionThe dilution solved all of the observed interferences. The results between diluted and manual counts showed a strong correlation, leading us to introduce dilution in our laboratory routine.