Pulse oximetry signal loss during hypoxic episodes in preterm infants receiving automated oxygen control.
Trial report in European journal of pediatrics, 2024. The graph could read no effect estimate from its abstract, so it casts no vote on the map. It is linked to trial NCT03785899 (Automatic Oxygen Control), which is not on this map. Not yet cited in PubMed.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Automatic Oxygen Control (SPOC) in Preterm Infants - Evaluation of a Revised Algorithm and Effect of Averaging Time of Pulse Oximetry Signal
Who cites it
0 citing papers in PubMed.
No citing paper in PubMed yet.
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
5 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
The aim of this study was to analyze signal loss (SL) resulting from low signal quality of pulse oximetry-derived hemoglobin oxygen saturation (SpO CLINICALTRIALS: gov under the registration no. NCT03785899. WHAT IS KNOWN: • Previous studies examined SpO2 signal loss (SL) during routine manual oxygen control being rare, but pronounced in lower SpO2 states. • Oxygen titration during SL is unlikely to be beneficial as SpO2 may recover to a normoxic range. WHAT IS NEW: • Periods of low signal quality of SpO2 are common during pHEs and while supported with automated oxygen control (SPOC), FiO2 is set to a back-up value reflecting FiO2 requirements during normoxemia in response to SL, although SpO2 remained below target until signal recovery. • FiO2 overshoots following pHEs were rare during AOC and occurred with a delayed onset; therefore, increased FiO2 during SL does not necessarily lead to overshoots.