Feasibility and Safety of Percutaneous Puncture Guided by a 5G-Based Telerobotic Ultrasound System: An Experimental Study.
Article in Cardiovascular and interventional radiology, 2024. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 1 paper.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
1 citing paper in PubMed.
- A shared-control ultrasound-guided robotic system for liver puncture under respiratory motion.Scientific reports · 2026Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
10 authors.
Funding
Abstract
purposeTo evaluate the feasibility and safety of percutaneous puncture guided by a 5th generation mobile communication technology (5G)-based telerobotic ultrasound system in phantom and animal experiments. MATERIALS AND
methodsIn the phantom experiment, 10 simulated lesions were punctured, once at each of two angles for each lesion, under the guidance of a telerobotic ultrasound system and ultrasound-guided freehand puncture. Student's t test was used to compare the two methods in terms of puncture accuracy, total operation duration, and puncture duration. In the animal experiment, under the guidance of the telerobotic ultrasound system, an 18G puncture needle was used to puncture 3 target steel beads in the liver, right kidney, and right gluteal muscle, respectively. The animal experiment had no freehand ultrasound-guided control group. After puncture, a CT scan was performed to verify the position of the puncture needle in relation to the target, and the complications and puncture duration, etc., were recorded.
resultsIn the phantom experiment, the mean accuracies of puncture under telerobotic ultrasound guidance and conventional ultrasound guidance were 1.8 ± 0.3 mm and 1.6 ± 0.3 mm (P = 0.09), respectively; therefore, there was no significant difference in the accuracy of the two guide methods. In the animal experiment, the first-attempt puncture success (the needle tip close to the target) rate was 93%. Polypnea occurred during one puncture. No other intraoperative or postoperative complications were observed.
conclusionPuncture guided by a 5G-based telerobotic ultrasound system has shown good feasibility and safety in phantom and animal experiments.