ArticleSICOT-J2024
Flexion contracture can cause component mismatch in the Prophecy
Article in SICOT-J, 2024. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Not yet cited in PubMed.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
0 citing papers in PubMed, 1 citations in OpenAlex.
No citing paper in PubMed yet.
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
4 authors at 1 institution in 1 country.
Funding
Abstract
introductionPatient-specific instrumentation (PSI) systems are used to conduct total knee arthroplasty. PSI reduces operative time, is less invasive and easier to use, and minimizes the risk of errors by providing precise measurements and reducing operating room turnover time. However, a study on the accuracy of Prophecy Evolution PSI (Microport Inc., Arlington, TN, USA) reported that 94% were below the error margin of 1.5 mm and 90% had error margins of 1 mm. This study aimed to evaluate the accuracy of the Prophecy Evolution PSI system in terms of the thickness of "total" bony resection required to achieve adequate extension/flexion gaps and the component match ratio between preoperative planning and actual component size inserted.
methodsComparisons were made between the sizes of femoral and tibial components planned with PSI and those inserted. The primary outcome was the average preoperative range of motion with and without matched femoral/tibial components. The study further analyzed the proportions of cases in which both the femoral and tibial components matched, neither matched, and only one of the femoral or tibial components matched.
resultsThe ratio of the same sizes between the PSI planning and those inserted was 50.8% (33 patients) for both the femoral and tibial components. For the femoral component alone, the ratio was 84.6% (55 patients), and for the tibial component, it was 58.4% (38 patients). A receiver-operating characteristic curve analysis indicated that flexion contracture greater than 20° was a significant prognostic factor for the PSI component match group versus the mismatch group. DISCUSSION: Flexion contracture may cause PSI mismatch. Notably, flexion contracture greater than 20° was a significant risk factor for the PSI component match group versus the mismatch group. During preoperative planning for a patient with flexion contracture, surgeons should prepare for the possibility of inserting an undersized tibial component.
Indexed as
Identifiers
What OpenQuestion holds
Registered trials