MR signature matching (MRSIGMA) implementation for true real-time free-breathing volumetric imaging with sub-200 ms latency on an MR-Linac.
Article in Magnetic resonance in medicine, 2024. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 3 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
3 citing papers in PubMed, 4 citations in OpenAlex.
- 3D and 4D Free-Breathing Abdominal T1-Weighted MRI in Clinical Practice Using Deep Learning Auto-Navigation and Reconstruction.Magnetic resonance in medicine · 2026Article
- Dynamic Mode Decomposition (DMD) for Low-Latency Real-Time Cardiac MRI.Magnetic resonance in medicine · 2026Article
- High-definition motion-resolved MRI using 3D radial kooshball acquisition and deep learning spatial-temporal 4D reconstruction.Physics in medicine and biology · 2025Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
8 authors at 2 institutions in 2 countries.
Funding
Abstract
purposeDevelop a true real-time implementation of MR signature matching (MRSIGMA) for free-breathing 3D MRI with sub-200 ms latency on the Elekta Unity 1.5T MR-Linac.
methodsMRSIGMA was implemented on an external computer with a network connection to the MR-Linac. Stack-of-stars with partial k
resultsMotion dictionary was computed in under 5 s. Signature acquisition and matching presented 173 ms latency on the phantom and 193 ms on patients. MRSIGMA presented a mean error of 1.3-1.6 mm for all phantom experiments, which was below the 2 mm acquisition resolution along the motion direction. The Dice coefficient over time between MRSIGMA and reference contours was 0.88 ± 0.02 (GTV), 0.87 ± 0.02(duodenum-stomach), and 0.78 ± 0.02(small bowel), demonstrating high motion tracking performance for both tumor and organs at risk.
conclusionThe real-time implementation of MRSIGMA enabled true real-time free-breathing 3D MRI with sub-200 ms imaging latency on a clinical MR-Linac system, which can be used for treatment monitoring, adaptive radiotherapy and dose accumulation mapping in tumors affected by respiratory motion.