ArticleAnnals of laboratory medicine2024

Integration of an MC-80 Digital Image Analyzer With an Automated BC-6800Plus Hematology Analyzer Enables Accurate Platelet Counting in Samples With EDTA-Induced Pseudothrombocytopenia.

Min-Kyung So, Jungwon Huh, Seunghwan Kim, Sholhui Park

Open access · diamondAbstract read
In one paragraph

Article in Annals of laboratory medicine, 2024. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 4 papers.

0numbers the graph read from it
0cells of the map it votes in
4citing papers in PubMed
1.0field-weighted citation impact, top 26% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

4 citing papers in PubMed, 3 citations in OpenAlex.

  1. Laboratory Assessment and Clinical Outlook forAnnals of laboratory medicine · 2026
    Review
  2. Review
  3. Article
  4. Review
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

4 authors at 1 institution in 1 country.

Min-Kyung SoDepartment of Laboratory Medicine, Ewha Womans University College of Medicine, Seoul, Korea.ORCID https://orcid.org/0000-0001-9728-3008
Jungwon HuhDepartment of Laboratory Medicine, Ewha Womans University College of Medicine, Seoul, Korea.ORCID https://orcid.org/0000-0001-8758-9847
Seunghwan KimDepartment of Laboratory Medicine, Ewha Womans University College of Medicine, Seoul, Korea.ORCID https://orcid.org/0000-0001-9923-4357
Sholhui ParkDepartment of Laboratory Medicine, Ewha Womans University College of Medicine, Seoul, Korea.ORCID https://orcid.org/0000-0002-6467-427X
Ewha Womans University · KR

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

Background: EDTA-induced pseudothrombocytopenia (PTCP) during whole blood collection requires significant laboratory resources to obtain accurate results. We evaluated platelet-deaggregation function in EDTA-induced PTCP and platelet-clump flagging by the BC-6800Plus hematology analyzer using integrated digital image analysis. Methods: We prospectively collected 132 whole blood samples suspected of platelet clumping (102 in EDTA and 30 in sodium citrate) from 88 individuals. We compared platelet counts determined using the platelet count by impedance (PLT-I) function of the DxH 900 hematology analyzer and the PLT-I or optical platelet count (PLT-O) function of the BC-6800Plus. Platelet clumping was verified through manual inspection and the MC-80 digital image analyzer. Results: Among the 132 whole blood samples, 43 EDTA samples showed platelet clumping. The DxH 900 PLT-I and BC-6800Plus PLT-I results demonstrated a strong correlation (r=0.711) for the EDTA samples but only a moderate correlation with the BC-6800Plus PLT-O results (r=0.506 and 0.545, respectively). The BC-6800Plus PLT-O results were consistent with the sodium citrate platelet counts, with a median dissociation rate of 102.5% (range, 74.9%-123.1%). The DxH 900 and BC-6800Plus analyzers had sensitivity values of 0.79 and 0.72, respectively, for platelet-clump flagging. When integrating the MC-80 digital image analysis results, the sensitivity of BC-6800Plus improved to 0.89 (standard mode) or 1.0 (PLT-Pro mode). Conclusions: BC-6800Plus PLT-O measurement results are close to the actual values obtained by platelet deaggregation with PTCP samples. Integrating the BC-6800Plus with a digital imaging analyzer effectively improved the diagnosis of PTCP and reduced the requirement for additional laboratory procedures.

Indexed as

Edetic AcidThrombocytopeniaBlood PlateletsFemaleHumansImage Processing, Computer-AssistedMalePlatelet AggregationPlatelet CountProspective StudiesSodium CitrateEdetic AcidSodium CitrateAnalyzerAnticoagulantComplete blood countDigital imageEDTAPseudothrombocytopenia

Identifiers

PMID38572537
PMCPMC11375204
OpenAlexW4393942959

What OpenQuestion holds

Textmetadata
LicenceCC BY-NC
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.