Evidence map›Paper›PMID 38572304›Full record

ArticleFrontiers in cardiovascular medicine2024

Diaphragm movement sensor for phrenic nerve monitoring during cryoballoon procedures: the first clinical evaluation.

Elsa Schemoul, Lilith Tovmassian, Julien Mancini, Linda Koutbi, Cédric Biermé, Jean-Claude Deharo, Frédéric Franceschi, Baptiste Maille

Abstract read
In one paragraph

Article in Frontiers in cardiovascular medicine, 2024. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 3 papers.

0numbers the graph read from it
0cells of the map it votes in
3citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

3 citing papers in PubMed.

  1. Trial
  2. Article
  3. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

8 authors.

Elsa SchemoulDepartment of Cardiology, CHU Timone, Aix-Marseille University, Marseille, France.
Lilith TovmassianDepartment of Cardiology, CHU Timone, Aix-Marseille University, Marseille, France.
Julien ManciniDepartment of Cardiology, CHU Timone, Aix-Marseille University, Marseille, France.
Linda KoutbiDepartment of Cardiology, CHU Timone, Aix-Marseille University, Marseille, France.
Cédric BierméDepartment of Cardiology, CHU Timone, Aix-Marseille University, Marseille, France.
Jean-Claude DeharoDepartment of Cardiology, CHU Timone, Aix-Marseille University, Marseille, France.
Frédéric FranceschiDepartment of Cardiology, CHU Timone, Aix-Marseille University, Marseille, France.
Baptiste MailleDepartment of Cardiology, CHU Timone, Aix-Marseille University, Marseille, France.

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

Background and aims: Right phrenic nerve palsy is the most frequent complication of cryoballoon procedures. The SMARTFREEZE™ console (Boston Scientific, St. Paul, MN, USA) has integrated a new tool for diaphragm monitoring-the Diaphragm Movement Sensor; however, it has not been evaluated in clinical practice. We aimed to assess the diagnostic performance of the Diaphragm Movement Sensor based on compound motor action potential data recorded simultaneously. Methods: Thirty consecutive patients (mean age 63.2 ± 10.2 years) were included. We simultaneously recorded the compound motor action potential and the Diaphragm Movement Sensor during cryoapplications in the right pulmonary veins. The right phrenic nerve was paced at 60 per minute, 12 V and 2.9 ms. Compound motor action potential monitoring with a 30% decrease cutoff for the diagnosis of phrenic nerve threatening was considered the gold standard. The Diaphragm Movement Sensor decrease threshold was also set at 30%. Results: Considering compound motor action potential monitoring, phrenic nerve threatening occurred 11 times (in seven patients) among 84 cryoapplications (13.1%) at the right pulmonary veins. The sensitivity and specificity of the Diaphragm Movement Sensor were, respectively, 33% (95% CI: 7%-70%) and 49% (95% CI: 38%-61%; Conclusions: The diagnostic performance of the Diaphragm Movement Sensor is low, and the relevance of its use in clinical practice may be debated.

Indexed as

atrial fbrillationCMAPcomplicationcryoballoon ablationphrenic nerve—injuries

Identifiers

PMID38572304
PMCPMC10987823

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.