Assessing the sensitivity and suitability of a range of detectors for SIMT PSQA.

Leon Dunn, Alessia Tamborriello, Brindha Subramanian, Xiaolei Xu, Tyrone Te Ruruku

Open access · goldAbstract read
In one paragraph

Article in Journal of applied clinical medical physics, 2024. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 9 papers.

0numbers the graph read from it
0cells of the map it votes in
9citing papers in PubMed
3.8field-weighted citation impact, top 6% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

9 citing papers in PubMed, 9 citations in OpenAlex.

  1. Article
  2. Article
  3. Article
  4. Article
  5. Review
  6. Article
  7. Article
  8. Article
  9. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

5 authors at 3 institutions in 2 countries.

Leon DunnGenesisCare Berwick, St John of God Berwick Specialist Centre, Berwick, Victoria, Australia.
Alessia TamborrielloGenesisCare Frankston, Frankston Public Surgical Centre, Frankston, Victoria, Australia.
Brindha SubramanianGenesisCare Ringwood, Ringwood Private Hospital, Ringwood East, Victoria, Australia.
Xiaolei XuGenesisCare Ringwood, Ringwood Private Hospital, Ringwood East, Victoria, Australia.
Tyrone Te RurukuGenesisCare Berwick, St John of God Berwick Specialist Centre, Berwick, Victoria, Australia.
Hollywood Private Hospital · AUR&G Associates · GBFrankston Hospital · AU

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

purposeSingle-isocenter multi-target intracranial stereotactic radiotherapy (SIMT) is an effective treatment for brain metastases with complex treatment plans and delivery optimization necessitating rigorous quality assurance. This work aims to assess five methods for quality assurance of SIMT treatment plans in terms of their suitability and sensitivity to delivery errors.

methodsSun Nuclear ArcCHECK and SRS MapCHECK, GafChromic EBT Radiochromic Film, machine log files, and Varian Portal Dosimetry were all used to measure 15 variations of a single SIMT plan. Variations of the original plan were created with Python. They comprised various degrees of systematic MLC offsets per leaf up to 2 mm, random per-leaf variations with differing minimum and maximum magnitudes, simulated collimator, and dose miscalibrations (MU scaling). The erroneous plans were re-imported into Eclipse and plan-quality degradation was assessed by comparing each plan variation to the original clinical plan in terms of the percentage of clinical goals passing relative to the original plan. Each erroneous plan could be then ranked by the plan-quality degradation percentage following recalculation in the TPS so that the effects of each variation could be correlated with γ pass rates and detector suitability. RESULTS &

conclusionsIt was found that 2%/1 mm is a good starting point for the ArcCHECK, Portal Dosimetry, and the SRS MapCHECK methods, respectively, and provides clinically relevant error detection sensitivity. Looser dose criteria of 5%/1 mm or 5%/1.5 mm are suitable for film dosimetry and log-file-based methods. The statistical methods explored can be expanded to other areas of patient-specific QA and detector assessment.

Indexed as

Brain NeoplasmsQuality Assurance, Health CareRadiosurgeryRadiotherapy DosageRadiotherapy, Intensity-ModulatedRadiotherapy Planning, Computer-AssistedAlgorithmsHumansParticle AcceleratorsRadiometryArcCHECKdetectorexternal beamfilmlog fileportal dosimetryPSQAquality assurancesensitivitySIMTSRSSRS MapCHECKSRTstereotactic

Identifiers

PMID38569013
PMCPMC11087180
OpenAlexW4393864578

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.