Evidence map›Paper›PMID 38561975›Full record

ArticleJournal of applied clinical medical physics2024

Repurposing DailyQA3 for an efficient and spot position sensitive daily quality assurance tool for proton therapy.

Hong Qi Tan, Calvin Wei Yang Koh, Kah Seng Lew, Ping Lin Yeap, Clifford Ghee Ann Chua, James Kuan Huei Lee, Yun Ming Wong, Andrew Wibawa, Zubin Master, James Cheow Lei Lee and 1 more

Abstract read
In one paragraph

Article in Journal of applied clinical medical physics, 2024. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 1 paper.

0numbers the graph read from it
0cells of the map it votes in
1citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

1 citing paper in PubMed.

  1. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

11 authors.

Hong Qi TanDivision of Radiation Oncology, National Cancer Centre Singapore, Singapore, Singapore.ORCID https://orcid.org/0000-0001-7878-4544
Calvin Wei Yang KohDivision of Radiation Oncology, National Cancer Centre Singapore, Singapore, Singapore.
Kah Seng LewDivision of Radiation Oncology, National Cancer Centre Singapore, Singapore, Singapore.
Ping Lin YeapDivision of Radiation Oncology, National Cancer Centre Singapore, Singapore, Singapore.
Clifford Ghee Ann ChuaDivision of Radiation Oncology, National Cancer Centre Singapore, Singapore, Singapore.
James Kuan Huei LeeDepartment of Physics, National University Singapore, Singapore, Singapore.
Yun Ming WongDivision of Physics and Applied Physics, Nanyang Technological University, Singapore, Singapore.
Andrew WibawaDivision of Radiation Oncology, National Cancer Centre Singapore, Singapore, Singapore.
Zubin MasterDivision of Radiation Oncology, National Cancer Centre Singapore, Singapore, Singapore.
James Cheow Lei LeeDivision of Radiation Oncology, National Cancer Centre Singapore, Singapore, Singapore.
Sung Yong ParkDivision of Radiation Oncology, National Cancer Centre Singapore, Singapore, Singapore.

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

introductionDaily quality assurance is an integral part of a radiotherapy workflow to ensure the dose is delivered safely and accurately to the patient. It is performed before the first treatment of the day and needs to be time and cost efficient for a multiple gantries proton center. In this study, we introduced an efficient method to perform QA for output constancy, range verification, spot positioning accuracy and imaging and proton beam isocenter coincidence with DailyQA3.

methodsA stepped acrylic block of specific dimensions is fabricated and placed on top of the DailyQA3 device. Treatment plans comprising of two different spread-out Bragg peaks and five individual spots of 1.0 MU each are designed to be delivered to the device. A mathematical framework to measure the 2D distance between the detectors and individual spot is introduced and play an important role in realizing the spot positioning and centering QA. Lastly, a 5 months trends of the QA for two gantries are presented.

resultsThe outputs are monitored by two ion chambers in the DailyQA3 and a tolerance of

conclusionWith the proposed approach, we can perform four QA items in the TG224 within 10 min.

Indexed as

Proton TherapyQuality Assurance, Health CareRadiotherapy DosageRadiotherapy Planning, Computer-AssistedAlgorithmsHumansNeoplasmsPhantoms, ImagingRadiometryRadiotherapy, Intensity-ModulatedDailyQA3pencil beam scanningproton therapyquality assurance

Identifiers

PMID38561975
PMCPMC11244688

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

None linked