A four-grating interferometer for x-ray multi-contrast imaging.
Article in Medical physics, 2024. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Not yet cited in PubMed.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
0 citing papers in PubMed.
No citing paper in PubMed yet.
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
3 authors.
Funding
Abstract
backgroundX-ray multi-contrast imaging with gratings provides a practical method to detect differential phase and dark-field contrast images in addition to the x-ray absorption image traditionally obtained in laboratory or hospital environments. Systems have been developed for preclinical applications in areas including breast imaging, lung imaging, rheumatoid arthritis hand imaging and kidney stone imaging. PURPOSE: Prevailing x-ray interferometers for multi-contrast imaging include Talbot-Lau interferometers and universal moiré effect-based phase-grating interferometers. Talbot-Lau interferometers suffer from conflict between high interferometer sensitivity and large field of view (FOV) of the object being imaged. A small period analyzer grating is necessary to simultaneously achieve high sensitivity and large FOV within a compact imaging system but is technically challenging to produce for high x-ray energies. Phase-grating interferometers suffer from an intrinsic fringe period ranging from a few micrometers to several hundred micrometers that can hardly be resolved by large area flat panel x-ray detectors. The purpose of this work is to introduce a four-grating x-ray interferometer that simultaneously allows high sensitivity and large FOV, without the need for a small period analyzer grating.
methodsThe four-grating interferometer consists of a source grating placed downstream of and close to the x-ray source, a pair of phase gratings separated by a fixed distance placed downstream of the source grating, and an analyzer grating placed upstream of and close to the x-ray detector. The object to be imaged is placed upstream of and close to the phase-grating pair. The distance between the source grating and the phase-grating pair is designed to be far larger than that between the phase-grating pair and the analyzer grating to promote simultaneously high sensitivity and large FOV. The method was evaluated by constructing a four-grating interferometer with an 8 µm period source grating, a pair of phase gratings of 2.4 µm period, and an 8 µm period analyzer grating.
resultsThe fringe visibility of the four-grating interferometer was measured to be ≈24% at 40 kV and ≈18% at 50 kV x-ray tube operating voltage. A quartz bead of 6 mm diameter was imaged to compare the theoretical and experimental phase contrast signal with good agreement. Kidney stone specimens were imaged to demonstrate the potential of such a system for classification of kidney stones.
conclusionsThe proposed four-grating interferometer geometry enables a compact x-ray multi-contrast imaging system with simultaneously high sensitivity and large FOV. Relaxation of the requirement for a small period analyzer grating makes it particularly suitable for high x-ray energy applications such as abdomen and chest imaging.