ArticleInternational journal of radiation oncology, biology, physics2024
Rapid Switching of a C-Series Linear Accelerator Between Conventional and Ultrahigh-Dose-Rate Research Mode With Beamline Modifications and Output Stabilization.
Article in International journal of radiation oncology, biology, physics, 2024. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 8 papers, 2 of them syntheses that pooled it.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
8 citing papers in PubMed, 2 syntheses or guidelines pooled it.
- A systematic review of electron FLASH dosimetry and beam control mechanisms utilized with modified non-clinical LINACs.Journal of applied clinical medical physics · 2025Pooled it
- The evolution of FLASH radiotherapy: a bibliometric analysis.Frontiers in oncology · 2025Pooled it
- Noninvasive Reversible Software-Based Electron FLASH Irradiation Configuration of a Linear Accelerator in Clinical Use.International journal of radiation oncology, biology, physics · 2026Article
- Development and evaluation of an in vivo dose-based monitoring system for electron FLASH radiation therapy.Medical physics · 2026Article
- Timescale of FLASH Sparing Effect Determined by Varying Temporal Split of Dose Delivery in Mice.International journal of radiation oncology, biology, physics · 2026Article
- Beam intensity and stability control on a modified clinical linear accelerator for FLASH irradiation.Physics in medicine and biology · 2025Article
- Article
- Mean dose rate in ultra-high dose rate electron irradiation is a significant predictor for OPhysics in medicine and biology · 2023Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
10 authors.
Funding
Abstract
purposeIn this study, a C-series linear accelerator was configured to enable rapid and reliable conversion between the production of conventional electron beams and an ultrahigh-dose-rate (UHDR) electron beamline to the treatment room isocenter for FLASH radiation therapy. Efforts to tune the beam resulted in a consistent, stable UHDR beamline. METHODS AND MATERIALS: The linear accelerator was configured to allow for efficient switching between conventional and modified electron output modes within 2 minutes. Additions to the air system allow for retraction of the x-ray target from the beamline when the 10 MV photon mode is selected. With the carousel set to an empty port, this grants access to the higher current pristine electron beam normally used to produce clinical photon fields. Monitoring signals related to the automatic frequency control system allows for tuning of the waveguide while the machine is in a hold state so a stable beam is produced from the initial pulse. A pulse counting system implemented on an field-programmable gate array-based controller platform controls the delivery to a desired number of pulses. Beam profiles were measured with Gafchromic film. Pulse-by-pulse dosimetry was measured using a custom electrometer designed around the EDGE diode.
resultsThis method reliably produces a stable UHDR electron beam. Open-field measurements of the 16-cm full-width, half-maximum gaussian beam saw average dose rates of 432 Gy/s at treatment isocenter. Pulse overshoots were limited and ramp up was eliminated. Over the last year, there have been no recorded incidents that resulted in machine downtime due to the UHDR conversions.
conclusionsStable 10 MeV UHDR beams were generated to produce an average dose rate of 432 Gy/s at the treatment room isocenter. With a reliable pulse-counting beam control system, consistent doses can be delivered for FLASH experiments with the ability to accommodate a wide range of field sizes, source-to-surface distances, and other experimental apparatus that may be relevant for future clinical translation.