ArticleJournal of applied clinical medical physics2024
Comprehensive beam delivery latency evaluation for gated proton therapy system using customized multi-channel signal acquisition platform.
Article in Journal of applied clinical medical physics, 2024. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 2 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
2 citing papers in PubMed.
- STAR in a Second: Ultra-High-Dose-Rate Spread-Out Bragg Peak Proton Therapy for Breath-Hold Stereotactic Arrhythmia Radioablation.Advances in radiation oncology · 2026Article
- Comprehensive beam delivery latency evaluation for gated proton therapy system using customized multi-channel signal acquisition platform.Journal of applied clinical medical physics · 2024Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
5 authors.
Funding
Abstract
purposeBeam delivery latency in respiratory-gated particle therapy systems is a crucial issue to dose delivery accuracy. The aim of this study is to develop a multi-channel signal acquisition platform for investigating gating latencies occurring within RPM respiratory gating system (Varian, USA) and ProBeam proton treatment system (Varian, USA) individually.
methodsThe multi-channel signal acquisition platform consisted of several electronic components, including a string position sensor for target motion detection, a photodiode for proton beam sensing, an interfacing board for accessing the trigger signal between the respiratory gating system and the proton treatment system, a signal acquisition device for sampling and synchronizing signals from the aforementioned components, and a laptop for controlling the signal acquisition device and data storage. RPM system latencies were determined by comparing the expected gating phases extracted from the motion signal with the trigger signal's state turning points. ProBeam system latencies were assessed by comparing the state turning points of the trigger signal with the beam signal. The total beam delivery latencies were calculated as the sum of delays in the respiratory gating system and the cyclotron proton treatment system. During latency measurements, simulated sinusoidal motion were applied at different amplitudes and periods for complete beam delivery latency evaluation under different breathing patterns. Each breathing pattern was repeated 30 times for statistical analysis.
resultsThe measured gating ON/OFF latencies in the RPM system were found to be 104.20 ± 13.64 ms and 113.60 ± 14.98 ms, respectively. The measured gating ON/OFF delays in the ProBeam system were 108.29 ± 0.85 ms and 1.20 ± 0.04 ms, respectively. The total beam ON/OFF latencies were determined to be 212.50 ± 13.64 ms and 114.80 ± 14.98 ms.
conclusionWith the developed multi-channel signal acquisition platform, it was able to investigate the gating lags happened in both the respiratory gating system and the proton treatment system. The resolution of the platform is enough to distinguish the delays at the millisecond time level. Both the respiratory gating system and the proton treatment system made contributions to gating latency. Both systems contributed nearly equally to the total beam ON latency, with approximately 100 ms. In contrast, the respiratory gating system was the dominant contributor to the total beam OFF latency.
Indexed as
Identifiers
What OpenQuestion holds
Registered trials
Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.