Repeatability of Pentacam-derived intraocular lens decentration measurements and the level of agreement with OPD-Scan III: A prospective observational case series.
Observational study in PloS one, 2024. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 1 paper.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
1 citing paper in PubMed.
- Tolerance to intraocular lens decentration and tilt in five presbyopia-correcting intraocular lenses: SS-AS-OCT-based study.Journal of cataract and refractive surgery · 2026Observational
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
6 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
purposeThis study aimed to assess the repeatability of intraocular lens (IOL) decentration measurements obtained through Pentacam, based on corneal topographic axis (CTA) and pupillary axis (PA), and to evaluate the level of agreement between Pentacam and OPD-Scan III devices in measuring IOL decentration.
methodsIn this prospective observational case series, three measurements were performed with Pentacam to evaluate the repeatability of the measurements. The analysis included the calculation of the mean and standard deviations (SD), conducting a repeated measures analysis of variance (rANOVA), and determining an intraclass correlation coefficient (ICC) to assess the repeatability of the measurements. Moreover, Bland-Altman analysis was employed to assess the agreement between Pentacam and OPD-Scan III devices in measuring IOL decentration. IOL decentration measurements were obtained with respect to both CTA and PA.
resultsA total of 40 eyes from 40 patients were analyzed. The rANOVA revealed no significant difference among three consecutive measurements of IOL decentration obtained with Pentacam. The mean SD of all parameters ranged from 0.04 mm to 0.07 mm. With CTA as the reference axis, the ICC values for Pentacam measurements of IOL decentration were 0.82 mm for the X-axis, 0.76 mm for the Y-axis, and 0.82 mm for spatial distance. When using PA as the reference axis, the corresponding ICC values were 0.87, 0.89, and 0.77, respectively. The 95% limits of agreement for all IOL decentration measurements were wide when comparing Pentacam and OPD-Scan III.
conclusionsPentacam demonstrated high repeatability in measuring IOL decentration with respect to both CTA and PA. However, due to poor agreement between Pentacam and OPD-Scan III measurements, caution should be exercised when using data interchangeably between the two devices.