ArticlePloS one2024

Accurate and reproducible enumeration of CD4 T cell counts and Hemoglobin levels using a point of care system: Comparison with conventional laboratory based testing systems in a clinical reference laboratory in Cameroon.

Bertrand Sagnia, Fabrice Mbakop Ghomsi, Sylvie Moudourou, Ana Gutierez, Jules Tchadji, Samuel Martin Sosso, Alexis Ndjolo, Vittorio Colizzi

Open access · goldAbstract read
In one paragraph

Article in PloS one, 2024. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 1 paper.

0numbers the graph read from it
0cells of the map it votes in
1citing papers in PubMed
0.3field-weighted citation impact, top 46% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

1 citing paper in PubMed, 1 citations in OpenAlex.

  1. Review
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

8 authors at 3 institutions in 2 countries.

Bertrand SagniaLaboratory of Microbiology and Immunology of CIRCB, Yaounde, Cameroon.ORCID 0000-0002-6707-8473
Fabrice Mbakop GhomsiCentre de Santé Catholique de Nkolondom, Yaounde, Cameroon.
Sylvie MoudourouMedical Analysis Laboratory of CIRCB, Yaounde, Cameroon.
Ana GutierezCatholic Hospital of BIKOP, Mbalmayo, Cameroon.
Jules TchadjiLaboratory of Microbiology and Immunology of CIRCB, Yaounde, Cameroon.
Samuel Martin SossoMedical Analysis Laboratory of CIRCB, Yaounde, Cameroon.
Alexis NdjoloLaboratory of Microbiology and Immunology of CIRCB, Yaounde, Cameroon.
Vittorio ColizziLaboratory of Immunology, Faculty of Sciences, University of Rome "Tor Vergata", Rome, Italy.
Centre International De Reference Chantal Biya · CMCatholic University of Cameroon · CMUniversity of Rome Tor Vergata · IT

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

backgroundMeasurements of CD4 T cells and hemoglobin (Hb) are conventionally used to determine the immunological state and disease progression for HIV-infected patients. We obtained a small lightweight point-of-care device, the BD FACSPrestoTM in order to demonstrate its ability to deliver CD4 and Hb analysis in comparison with two larger clinical machines the BDFACSCantoTM analyzer and Sysmex XN 1000 haematology analyzer. The advantages of using the POC device include access to HIV patient data in remote and in resource limited settings.

methodThe analytical performance of the BD FACSPrestoTM, compared with the FACSCantoTM II flow cytometer and the Sysmex XN 1000 haematology analyzer was evaluated by testing 241 routine clinical specimens collected in EDTA tubes from patients attending the Immunology and Microbiology laboratory of Chantal BIYA International Reference Centre (Yaounde, Cameroon) between January and May 2016.

resultsThe mean in absolute counts and percentage of CD4 T cells was 606 cells/mL and 25% respectively via the FACSPrestoTM, and 574 cells/mL and 24% respectively via the BD FACSCantoTM II. The mean concentration of Hb levels was 11.90 on the Sysmex XN 1000 and 11.45 via the BD FACSPrestoTM, A high correlation (R2 = 0.95, P < 0.001) of Hb level measurements was noted between the BD FACSPrestoTM and Sysmex XN 1000 hematology analyzer. Overall, a Bland-Altman plot of the differences between the two methods showed an excellent agreement for absolute and percentage CD4 counts and hemoglobin measurements between POC and conventional methods evaluated here. Furthermore, the study demonstrated the ease of use of the BD FACSPrestoTM POC technology in remote areas.

conclusionThe BD FACPrestoTM is a suitable tool for CD4 enumeration in resource-limited settings, specifically providing a deployable, reliable POC testing option. The BD FACSPrestoTM performed appropriately in comparison to the conventional reference standard technologies. The BD FACSPrestoTM, system provides accurate, reliable, precise CD4/%CD4/Hb results on venous blood sampling. The data showed good agreement between the BD FACSPrestoTM, BD FACSCantoTM II and Sysmex XN 1000 XN 1000 systems.

Indexed as

CD4-Positive T-LymphocytesHIV InfectionsCameroonCD4 Lymphocyte CountCell CountHemoglobinsHumansPoint-of-Care SystemsReproducibility of ResultsHemoglobins

Identifiers

PMID38507344
PMCPMC10954178
OpenAlexW4392985474