ArticleEJNMMI physics2024
Comparison of a 3D CZT and conventional SPECT/CT system for quantitative Lu-177 SPECT imaging.
Article in EJNMMI physics, 2024. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 14 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
14 citing papers in PubMed, 16 citations in OpenAlex.
- State-of-the-art and future developments in dosimetry for SSTR- and PSMA-targeting radioligand therapies.Zeitschrift fur medizinische Physik · 2026Review
- Quantitative accuracy ofEJNMMI physics · 2026Article
- QuantitativeEJNMMI physics · 2026Article
- Multicenter comparison of radionuclide calibrators and SPECT/CT protocols for quantitativeEJNMMI physics · 2026Article
- 3D-ring CZT gamma cameras in theranostics: is it a revolution?EJNMMI reports · 2026Article
- A comparison of three-dimensional-ring CZT and NaI(Tl) single-photon emission computed tomography systems usingQuantitative imaging in medicine and surgery · 2026Article
- Fast [EJNMMI physics · 2026Article
- Inter-institutional variability in kidney dosimetry duringRadiological physics and technology · 2026Article
- Quantitative performance of a 3D CZT and conventional SPECT/CT for Tb-161 imaging.EJNMMI physics · 2026Article
- Recommendations for Standardizing Nuclear Medicine Terminology and Data in the Era of Theranostics and Artificial Intelligence.Journal of nuclear medicine : official publication, Society of Nuclear Medicine · 2025Article
- Letter to the Editor: EANM position paper on challenges and opportunities of full-ring 360° CZT bone imaging: it's time to let go of planar whole-body bone imaging.European journal of nuclear medicine and molecular imaging · 2025Article
- Bias and precision of SPECT-basedEJNMMI physics · 2024Article
- 360° CZT-SPECT/CT cameras:EJNMMI physics · 2024Article
- From SPECT/CT towards absolute quantification? - the case of unilateral condylar hyperplasia of the mandible.EJNMMI physics · 2024Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
4 authors at 3 institutions in 1 country.
Funding
No grant is acknowledged in the PubMed record.
Abstract
purposeNext-generation SPECT/CT systems with CdZnTe (CZT) digital detectors in a ring-like setup are emerging to perform quantitative Lu-177 SPECT imaging in clinical routine. It is essential to assess how the shorter acquisition time might affect the image quality and uncertainty on the mean absorbed dose of the tumors and organs at risk compared to a conventional system.
methodsA NEMA Image Quality phantom was scanned with a 3D CZT SPECT/CT system (Veriton, by Spectrum Dynamics) using 6 min per bed position and with a conventional SPECT/CT system (Symbia T16, by Siemens) using 16 min per bed position. The sphere-to-background ratio was 12:1 and the background activity concentration ranged from 0.52 to 0.06 MBq/mL. A clinical reconstruction protocol for dosimetry purposes was determined for both systems by maximizing the sphere-to-background ratio while keeping the coefficient of variation of the background as low as possible. The corresponding image resolution was determined by the matching filter method and used for a dose uncertainty assessment of both systems following an established uncertainty model..
resultsThe optimized iterative reconstruction protocol included scatter and attenuation correction for both systems and detector response modeling for the Siemens system. For the 3D CZT system, 6 iterations and 8 subsets were combined with a Gaussian post-filter of 3 mm Full Width Half Maximum (FWHM) for post-smoothing. For the conventional system, 16 iterations and 16 subsets were applied with a Gaussian post-smoothing filter of 1 mm FWHM. For these protocols, the sphere-to-background ratio was 18.5% closer to the true ratio for the conventional system compared to the 3D CZT system when considering the four largest spheres. Meanwhile, the background coefficient of variation was very similar for both systems. These protocols resulted in SPECT image resolution of 14.8 mm and 13.6 mm for the 3D CZT and conventional system respectively. Based on these resolution estimates, a 50% dose uncertainty corresponded to a lesion volume of 28 mL for the conventional system and a lesion volume of 33 mL for the 3D CZT system.
conclusionsAn optimized reconstruction protocol for a Veriton system with 6 min of acquisition time per bed position resulted in slightly higher dose uncertainties than a conventional Symbia system using 16 min of acquisition time per bed position. Therefore, a 3D CZT SPECT/CT allows to significantly reduce the acquisition times with only a very limited impact on dose uncertainties such that quantitative Lu-177 SPECT/CT imaging becomes much more accessible for treatment concurrent dosimetry. Nevertheless, the uncertainty of SPECT-based dose estimates remains high.