Performance analysis of urine formed element analyzer EH-2090 was found to have good accuracy in detecting RBCs and WBCs when compared to manual microscopic.

Lixia Bai, Qiaoping Xu, Zhicheng Wu

Open access · diamondAbstract read
In one paragraph

Article in Translational andrology and urology, 2024. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 3 papers.

0numbers the graph read from it
0cells of the map it votes in
3citing papers in PubMed
2.7field-weighted citation impact, top 10% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

3 citing papers in PubMed, 6 citations in OpenAlex.

  1. Article
  2. Article
  3. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

3 authors at 1 institution in 1 country.

Lixia Bai *Department of Laboratory Medicine, Peking University Shenzhen Hospital, Shenzhen, China.
Qiaoping Xu *Department of Laboratory Medicine, Peking University Shenzhen Hospital, Shenzhen, China.
Zhicheng WuDepartment of Laboratory Medicine, Peking University Shenzhen Hospital, Shenzhen, China.
Peking University Shenzhen Hospital · CN

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

Background: EH-2090 is Mindray's new-generation fully automatic urine formed element analyzer (hereinafter referred to as EH-2090). Currently, there are no studies on EH-2090, so we evaluated the analytical and clinical performance of this instrument to verify that it can meet daily clinical needs, and used manual microscopy as a reference method. Methods: The analytical performance of the EH-2090 was first evaluated for repeatability, linearity, reproducibility, and carryover. We collected urine samples from outpatient and inpatient departments of Peking University Shenzhen Hospital. Uncentrifuged urine was compared with the EH-2090 using the Fuchs-Rosenthal counting method-a quantitative reference method for microscopy-for comparative studies in terms of red blood cell (RBC) and white blood cell (WBC) counting accuracy. Passing-Bablok regression analysis was performed for RBC and WBC counts. Two laboratory technicians performed centrifugation and manual analysis (microscopy) to evaluate its performance at detecting RBCs, WBCs, and casts, sensitivities and specificities were calculated. Results: The EH-2090's between-run reproducibility, within-day reproducibility, between-day reproducibility, and within-laboratory reproducibility for formed components of urine all met the laboratory requirements. There was a good correlation between the counting accuracy of RBCs (r=0.965, P<0.0001) and WBCs (r=0.894, P<0.0001) by the EH-2090 and the Fuchs-Rosenthal method. The positive coincidence rates of RBC and manual microscopy were 86.08% and 92.41%, respectively, and the negative coincidence rates were 88.39% and 85.81%, respectively. The positive coincidence rates before and after the WBC review were 89.33% and 92.00%, respectively, whereas the negative ones were 77.64% and 83.23%, respectively. The positive coincidence rates before and after cast review were 77.78% and 82.05%, respectively, and the negative ones were 97.09% and 93.60%, respectively. Conclusions: The EH-2090 has good analytical and clinical performance. Its RBC and WBC counting accuracy correlates well with the quantitative reference method of microscopy.

Indexed as

manual microscopyMindray EH-2090Urine formed element analyzer

Identifiers

PMID38481871
PMCPMC10932639
OpenAlexW4392105988

What OpenQuestion holds

Textmetadata
LicenceCC BY-NC-ND
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.