Evidence map›Paper›PMID 38473684›Full record

ArticleMaterials (Basel, Switzerland)2024

Finite Element Analysis and Fatigue Test of INTEGRA Dental Implant System.

Rafał Zieliński, Sebastian Lipa, Martyna Piechaczek, Jerzy Sowiński, Agata Kołkowska, Wojciech Simka

Open access · goldAbstract read
In one paragraph

Article in Materials (Basel, Switzerland), 2024. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 5 papers.

0numbers the graph read from it
0cells of the map it votes in
5citing papers in PubMed
5.4field-weighted citation impact, top 4% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

5 citing papers in PubMed, 11 citations in OpenAlex.

  1. Review
  2. Article
  3. Review
  4. Article
  5. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

6 authors at 2 institutions in 1 country.

Rafał ZielińskiStomatologia na Ksiezym Mlynie, 16 D Tymienieckiego, 90-365 Lodz, Poland.ORCID 0000-0002-9897-5003
Sebastian LipaInstitute of Materials Science and Engineering, Lodz University of Technology, 90-924 Lodz, Poland.
Martyna PiechaczekStomatologia na Ksiezym Mlynie, 16 D Tymienieckiego, 90-365 Lodz, Poland.
Jerzy SowińskiPrivate Dental Clinic, Tetmajera 3A Rd., 05-080 Izabelin C, Poland.
Agata KołkowskaChemistry Students Research Society, Faculty of Chemistry, Silesian University of Technology, 44-100 Gliwice, Poland.
Wojciech SimkaChemistry Students Research Society, Faculty of Chemistry, Silesian University of Technology, 44-100 Gliwice, Poland.ORCID 0000-0002-2648-5523
Silesian University of Technology · PLLodz University of Technology · PL

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

The study involved numerical FEA (finite element analysis) of dental implants. Based on this, fatigue tests were conducted according to the PN-EN 14801 standard required for the certification of dental products. Thanks to the research methodology developed by the authors, it was possible to conduct a thorough analysis of the impact of external and internal factors such as material, geometry, loading, and assembly of the dental system on the achieved value of fatigue strength limit in the examined object. For this purpose, FEM studies were based on identifying potential sites of fatigue crack initiation in reference to the results of the test conducted on a real model. The actions described in the study helped in the final evaluation of the dental system design process named by the manufacturer as INTEGRA OPTIMA 3.35. The objective of the research was to identify potential sites for fatigue crack initiation in a selected dental system built on the INTEGRA OPTIMA 3.35 set. The material used in the research was titanium grade 4. A map of reduced von Mises stresses was used to search for potential fatigue crack areas. The research [loading] was conducted on two mutually perpendicular planes positioned in such a way that the edge intersecting the planes coincided with the axis of the system. The research indicated that the connecting screw showed the least sensitivity (stress change) to the change in the loading plane, while the value of preload has a significant impact on the achieved fatigue strength of the system. In contrast, the endosteal implant (root) and the prosthetic connector showed the greatest sensitivity to the change in the loading plane. The method of mounting [securing] the endosteal implant using a holder, despite meeting the standards, may contribute to generating excessive stress concentration in the threaded part. Observation of the prosthetic connector in the Optima 3.35 system, cyclically loaded with a force of F ≈ 300 N in the area of the upper hexagonal peg, revealed a fatigue fracture. The observed change in stress peak in the dental connector for two different force application surfaces shows that the positioning of the dental system (setting of the socket in relation to the force action plane) is significantly decisive in estimating the limited fatigue strength.

Indexed as

dental implant cracksdental implant damagefinite element analysisINTEGRA IMPLANTSINTEGRA OPTIMAPN-EN 14801

Identifiers

PMID38473684
PMCPMC10934518
OpenAlexW4392517239

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

None linked