Implantation accuracy of novel polyimide stereotactic electroencephalographic depth electrodes-a human cadaveric study.

Aura Kullmann, Farida Akberali, Jaime J Van Gompel, Robert A McGovern, W Richard Marsh, Debra Kridner, Camilo A Diaz-Botia, Michael C Park

Open access · goldAbstract read
In one paragraph

Article in Frontiers in medical technology, 2024. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 1 paper.

0numbers the graph read from it
0cells of the map it votes in
1citing papers in PubMed
0.8field-weighted citation impact, top 30% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

1 citing paper in PubMed, 2 citations in OpenAlex.

  1. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

8 authors at 3 institutions in 1 country.

Aura KullmannNeuroOne Medical Technologies, Eden Prairie, MN, United States.
Farida AkberaliZimmer Biomet, Warsaw, IN, United States.
Jaime J Van GompelDepartment of Neurosurgery, Mayo Clinic, Rochester, MN, United States.
Robert A McGovernDepartment of Neurosurgery, University of Minnesota Medical Center, Minneapolis, MN, United States.
W Richard MarshDepartment of Neurosurgery, Mayo Clinic, Rochester, MN, United States.
Debra KridnerNeuroOne Medical Technologies, Eden Prairie, MN, United States.
Camilo A Diaz-BotiaNeuroOne Medical Technologies, Eden Prairie, MN, United States.
Michael C ParkDepartment of Neurosurgery, University of Minnesota Medical Center, Minneapolis, MN, United States.
Mayo Clinic in Arizona · USUniversity of Minnesota Medical Center · USZimmer Biomet (United States) · US

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

Introduction: Stereoelectroencephalography (sEEG) is a minimally invasive procedure that uses depth electrodes stereotactically implanted into brain structures to map the origin and propagation of seizures in epileptic patients. Implantation accuracy of sEEG electrodes plays a critical role in the safety and efficacy of the procedure. This study used human cadaver heads, simulating clinical practice, to evaluate (1) neurosurgeon's ability to implant a new thin-film polyimide sEEG electrode according to the instructions for use (IFU), and (2) implantation accuracy. Methods: Four neurosurgeons (users) implanted 24 sEEG electrodes into two cadaver heads with the aid of the ROSA robotic system. Usability was evaluated using a questionnaire that assessed completion of all procedure steps per IFU and user errors. For implantation accuracy evaluation, planned electrode trajectories were compared with post-implantation trajectories after fusion of pre- and postoperative computer tomography (CT) images. Implantation accuracy was quantified using the Euclidean distance for entry point error (EPE) and target point error (TPE). Results: All sEEG electrodes were successfully placed following the IFU without user errors, and post-implant survey of users showed favorable handling characteristics. The EPE was 1.28 ± 0.86 mm and TPE was 1.61 ± 0.89 mm. Long trajectories (>50 mm) had significantly larger EPEs and TPEs than short trajectories (<50 mm), and no differences were found between orthogonal and oblique trajectories. Accuracies were similar or superior to those reported in the literature when using similar experimental conditions, and in the same range as those reported in patients. Discussion: The results demonstrate that newly developed polyimide sEEG electrodes can be implanted as accurately as similar devices in the marker without user errors when following the IFU in a simulated clinical environment. The human cadaver ex-vivo test system provided a realistic test system, owing to the size, anatomy and similarity of tissue composition to that of the live human brain.

Indexed as

epilepsyex-vivo modelsimplantation accuracysEEGsEEG-guided radiofrequency ablationstereotactic implantationusability

Identifiers

PMID38456122
PMCPMC10917981
OpenAlexW4392055888

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.